OIML R107 (Discontinuous Totalizing Automatic Weighing Instruments) Certificate Schema. The certificate structure lists the top-level fields of the document model. The definitions expand the family's own vocabulary. Where the corpus carries fill statistics, the observed fields table shows how often each characteristic actually appears.
Certificate structure
Every R107 certificate is one object with these top-level fields. A "core" shape is shared across all Recommendation families; the rest are defined by this family's own definitions below.
Field
Shape
Required
certificate
core · Certificate
yes
issuing_authority
core · IssuingAuthority
no
applicants
core · PartyList
no
manufacturers
core · PartyList
no
certified_type
CertifiedType
yes
characteristics
Characteristics
yes
recommendation
object
yes
test_reports
core · TestReportList
no
revision_history
core · RevisionHistory
no
model_family
core · ModelFamily
no
components
array of core · Component
no
footnotes
core · FootnoteList
no
matrix_tables
MatrixTableList
no
d011_tests_applicable
array of D011TestReference
no
Observed fields
From 3 digitalized R107 certificates: how often each type-level characteristic appears (fill rate), in how many certificates, and the most common values observed.
Field
Fill rate
Present in
Common values
temperature_range
67%
2
min=-10, max=40
max_intervals_n
33%
1
6000
max_intervals_n_partial
33%
1
3000
max_capacity_range
33%
1
min=3, max=300000
weighing_range
33%
1
Single-interval
number_of_scale_intervals
33%
1
min=None, max=10000
verification_scale_interval_dt
33%
1
min=1, max=None
verification_scale_interval_min_uv
33%
1
min=0.4, max=None
maximum_capacity
33%
1
n×dt
minimum_capacity
33%
1
≥20% of Max
minimum_totalized_load
33%
1
according to OIML R107-1 section 2.5
subtractive_tare
33%
1
≤-Max
power_supply_voltage
33%
1
5
load_cell_impedance
33%
1
min=43, max=1100
load_cell_connecting_system
33%
1
4-wire or 6-wire, shielded
module_fractional_factor_pi
33%
1
0.5
interface
33%
1
Protective, according to paragraph 5.3.6
connected_load_cells
33%
1
Shall comply with R60
supply_voltage
33%
1
min=12, max=28
weighing_mode
33%
1
Static
electromagnetic_environment_class
33%
1
E2
humidity_class
33%
1
Non-condensing
warm_up_time
33%
1
Not needed
software_identification
33%
1
2.xx
max_cable_length_to_junction_box
33%
1
4824
rate_of_operation
33%
1
determined at initial verification
number_of_scale_intervals_n
33%
1
6000