Skip to content

← R85

R85/2008-A-NL1-20.04 Rev. 3

Applicant
Enraf B.V.
Issuer
NL1
Issuing year
2025
Status
Valid

Source PDF (certificates/R85/2008/r85-2008-nl1-2020-04-rev3.pdf) · extracted by OCR from the digitalized tier of the corpus.

R85/2008-A-NL1-20.04 Rev. 3

OIML Certificate

OIML Member State The Netherlands

Number R85/2008-A-NL1-20.04 revision 3 Project number 3768512 Page 1 of 6

Issuing authority Person responsible:

NMi Certin B.V.

M.Ph.D. Schmidt

Applicant and Manufacturer

Enraf B.V.

Delftechpark 39

2628 XJ Delft

The Netherlands

Identification of the

An automatic level gauge (ALG)

certified type

Type: SmartRadar FlexLine XP and SmartRadar FlexLine HP with the antennas F06, F08, W06, H04, S06, S08, S10 and S12 with indicating device SmartView, and / or indicating device HART SmartView with field interface 880 CIU-Prime and / or 880 CIU-Plus and / or CIU 880 with remote calculating and indicating system EntisPro with indicating and memory device ENTIS

Characteristics See page 2 and further

This OIML Certificate is issued under scheme A.

This Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML Type Evaluation Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

R 85-1 & 2 (2008) "Automatic level gauges for measuring the level of liquid in stationary storage tanks"

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation identified above. This Certificate does not bestow any form of legal international approval.

This certificate and supporting reports comply with the requirements of OIML-CS-PD-07 clause 6.2.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Type Evaluation Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 14 March 2025

NMi Certin B.V.

Thijsseweg 11

2629 JA Delft

The Netherlands

T +31 88 636 2332

certin@nmi.nl

www.nmi.nl

Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

This document is digitally signed and sealed. The digital signature can be verified in the blue ribbon at the top of the electronic version of this certificate.

OIML Certificate

Number R85/2008-A-NL1-20.04 revision 3 Project number 3768512

Page 2 of 6

The conformity was established by the results of tests and examinations provided in the associated reports:

  • R85/1998-NL1-07.02 that includes 100 pages;

  • CPC/9200376 that includes 20 pages;

  • NMi-10200994 that includes 15 pages;

  • NMi-12200691 that includes 13 pages;

  • NMi-13200623 that includes 14 pages;

  • NMi-14200253-1 that includes 21 pages;

  • NMi-16200400-01 that includes 21 pages;

  • NMi-16200400-02 that includes 21 pages;

  • R85-2008-NL1-12.04 dated 10 December 2012 that includes 49 pages;

  • NMi-13200623 dated 15 October 2013 that includes 14 pages;

  • NMi-1900750-02 dated 24 March 2017 that includes 26 pages.

  • NMi-2391640-01 dated 25 September 2019 that includes 30 pages;

  • NMi-2406967-01 dated 7 November 2019 that includes 27 pages;

  • NMi-2458123-01 dated 26 April 2021 that includes 17 pages.

The automatic level gauge (ALG) is produced at the following production locations:

  • Enraf B.V., Delftechpark 39, 2628 XJ Delft, The Netherlands.

  • Honeywell International (India) Private Ltd., Plot No.2, Gat No.181, Village Fulgaon, Tal-Haveli, PUNE-412216, India

Characteristics of the measuring instrument

In Table 1 the general characteristics of the measuring instrument are presented.

The construction of the measuring instrument is recorded in the Documentation folder no. T7316-10.

Table 1 General characteristics

Measuring rangeSee table 2
Ambient temperature range-25-+70℃; condensing humidity
Power supply voltage65-264Vac(-15%/+10%);50/60Hz24-65Vdc
Software identificationSee table 4

Table 2 General characteristics of the family of instruments

TypeRangeMinimum and maximum values for liquid pressure, for liquid temperature and for liquid properties.Minimum and maximum values for vapour pressure, for vapour temperature and for vapour properties.
Stilling well35mThe manufacturer shall specify these values for each application.The manufacturer shall specify these values for each application.

OIML Certificate

OIML Member State The Netherlands

Number R85/2008-A-NL1-20.04 revision 3

Project number 3768512

Page 3 of 6

TypeRangeMinimum and maximum values for liquid pressure, for liquid temperature and for liquid properties.Minimum and maximum values for vapour pressure, for vapour temperature and for vapour properties.
Free space35mThe manufacturer shall specify these values for each application.The manufacturer shall specify these values for each application.

Table 3 Antennas

TypeApplicationDocumentation
F06free space7314/7-01
F08free space7314/5-24
W06free space7314/5-25
S06on a stilling well with an inner diameter of 6"7314/5-26
S08on a stilling well with an inner diameter of 8"7314/5-26
S10on a stilling well with an inner diameter of 10"7314/5-26
S12on a stilling well with an inner diameter of 12"7314/5-26
H04on a stilling well with an inner diameter of 4"7314/5-27

Table 4 Software identification

PartTypeVersionChecksum
sensor processor in combination with sensorART2ATII-XR(also indicated asCAN Xband board)withART2AA10xxx0
DSP A10 xxx
A11xxx
DSP A11 xxx
A12xxx
DSP A12 xxx
A130038676
DSP A1300
A130111461
DSP A1301
A130260856
DSP A1302

OIML Member State The Netherlands

OIML Certificate

Number R85/2008-A-NL1-20.04 revision 3

Project number 3768512

Page 4 of 6

PartTypeVersionChecksum
sensor processor in combination with sensorART2BTII-XR(also indicated asCAN Xband board)withART2BA10xxx0
DSP A10 xxx
A11xxx
DSP A11 xxx
A120464095(=0xFA5F)
DSP A12 xxx
A130038676
DSP A1300
A130111461
DSP A1301
A130260856
DSP A1302
display communicationboardHMI-TSI/FII-SMVA10xxx(up toA1006)0
A100603170(=0x0C62)
A100722441
FCI-HRTA1006T38785(=0x9781)
A100712537
A100854556
A100926293
A101049336
A101131984
A101256965
communicationboardCAN-BPM/HCI-BPMA10xxx(up toA1007)0
A100737556(=0x92B4)
A20003260
A20010243
A200350556
A200457365

OIML Certificate

OIML Member State The Netherlands

Number R85/2008-A-NL1-20.04 revision 3 Project number 3768512

Page 5 of 6

PartTypeVersionChecksum
communication boardCAN-TRL2/HCI-TRL2A100112361030(=0x00BC9D46)
A10123112553898(=0xB985CDAA)
A200034966
A200133304
interface boardCAN-RS/HCI-GPUA10xxx0
1 WL main boardCAN-HCI-1WLA10xxx0
A301322685
A301716395
A301811607
A302037576
A30219433
RTD boardFII-RTDA100624904

Certificate history:

This revision replaces the previous version.

RevisionDateDescription of the modification
Initial15 February 2021-
18 June 2021Addition of software version and production location
230 August 2021Addition of software version
314 March 2025Extension of the measuring range up to 35mAddition of RTD board, software version and checksumUpdate the display communication board software version and checksum

OIML Certificate

OIML Member State The Netherlands

Number R85/2008-A-NL1-20.04 revision 3 Project number 3768512 Page 6 of 6

Design of a Tank Gauging System - Overview