Skip to content

← R76

R76/2006-A-NL1-24.25 Rev. 0

Applicant
A&D Company, Limited
Issuer
NL1
Issuing year
2024
Status
Valid

Source PDF (certificates/R76/2006/r76-2006-nl1-2024-25-rev0.pdf) · extracted by OCR from the digitalized tier of the corpus.

R76/2006-A-NL1-24.25 Rev. 0

OIML Certificate

OIML Member State The Netherlands

Number R76/2006-A-NL1-24.25 revision 0

Project number 3834276

Page 1 of 2

Issuing authority

NMi Certin B.V. Person responsible: M.Ph.D. Schmidt

Applicant and Manufacturer

A&D Company, Limited 1-243 Asahi Kitamoto-shi, Saitama-ken Japan

Identification of the certified type

An Indicator Type

: AD-4421

Characteristics

This OIML Certificate is issued under scheme A.

This Certificate attests the conformity of the above identified Type (represented by the sample(s) identified in the OIML Test Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

See next page

OIML R 76-1:2006 for accuracy class or

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation above-identified. This Certificate does not bestow any form of legal international approval.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 23 December 2024

Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

OIML Certificate

OIML Member State The Netherlands

Number R76/2006-A-NL1-24.25 revision 0 Project number 3834276

Page 2 of 2

The conformity was established by the results of tests and examinations provided in the associated reports:

No. NMi-3834276-01 dated 23 December 2024 that includes 39 pages;

No. NMi-3834276-02 dated 23 December 2024 that includes 21 pages.

No. NMi-3834276-03 dated 23 December 2024 that includes 11 pages.

  • No. NMi-3834276-04 dated 23 December 2024 that includes 11 pages.

  • No. NMi-3834276-05 dated 23 December 2024 that includes 13 pages.

Characteristics of the indicator:

Analog load cells
Accuracy class OIML R 76or
Weighing rangeSingle interval
Maximum number of scale intervals(one weighing range)n≤10000 divisions
Load cell excitation voltage5V DC
Minimum input voltage per verification scale interval0.5μV
Minimum load cell resistance42Ω
Maximum load cell resistance1000Ω
Fraction of the maximum permissible error0.5
Load cell interface6-wire with sense technology, may be configured as 4-wire
Maximum value of the cable length per cross wire section between the indicator and the junction box or load cellsNo special cable lengthIn case sense technology is not used the load cells are connected directly without junction box or extension cable
Temperature range-10℃/+40℃
Power supply voltage100-240V AC 50/60Hz
Software identificationVersion number:1.x.xx(x.xx is a number between0.00 and9.99 and represents the non-legally relevant software)

Revision History

RevisionDateChanges
02024-12-23Initial issue.