R76/2006-A-CZ1-26.01
| OIML Certification System | CMI | V3112 |
| OIML Member State Czech Republic | OIML Certificate No. R76/2006-A-CZ1-26.01 | |
| OIML CERTIFICATE ISSUED UNDER SCHEME A | ||
| OIML Issuing Authority Name: Czech Metrology Institute Address: Okružní 31 638 00 Brno Czech Republic Person responsible: Jan Kalandra | ||
| Applicant Name: ADE Germany GmbH Address: Neuer Höltigbaum 15 22143 Hamburg Germany | ||
| Manufacturer Name: ADE Germany GmbH Address: Neuer Höltigbaum 15 22143 Hamburg Germany | ||
| Identification of the certified type (the detailed characteristics will be defined in the additional pages) Indicator for non-automatic weighing instruments. type: MZ3042 | ||
| Designation of the module (if applicable) | ||
| This OIML Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML type evaluation report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML): OIML R 76 Edition (year): 2006 For accuracy class: III | ||
| This OIML Certificate relates only to metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML Recommendation identified above. This OIML Certificate does not bestow any form of legal international approval. | ||
| The conformity was established by the results of tests and examinations provided in the associated. Test report No. 6012-PT-R0046-25 dated 09.01.2026 that includes 51 pages. Test report No. 6012-PT-R0048-25 dated 13.01.2026 that includes 17 pages. Test report No. 6012-PT-R0049-25 dated 09.01.2026 that includes 36 pages. Test report No. 8551-PT-E0338-25 dated 12.11.2025 that includes 37 pages. OIML type evaluation report: No. 0511-ER-N021-25 dated 14 January 2026 that includes 12 pages | ||
| The technical documentation relating to the identified type is contained in documentation file: 0511-UL-N021-25 OIML Certificate History | ||
| Revision No. | Date | Description of the modification |
| - | 23 January 2026 | Issuing of certificate |
| The OIML Issuing Authority Ing. František Staněk, PhD. Deputy Director of Certification Body Date: 23 January 2026 | ||
| Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate is issued, partial quotation of the Certificate and of the associated OIML type evaluation report(s) is not permitted, although either may be reproduced in full. | ||
1 Characteristics of the instrument
The electronic indicator type MZ3042 is intended for use as indicator for non-automatic weighing instruments for medical applications. The instrument is a Class III, self-indicating weighing instrument with single- or dual-range, an external AC mains adapter and an internal rechargeable battery (optional).
ANSICHT1 ( 1:1 )
ANSICHT 2 ( 1:2 )
| ADE | Gewicht 0,54 kg | 12 | |||
| ATT Nr. GmbH GmbH GmbH | |||||
| Att Nr. GmbH GmbH GmbH | |||||
| Datum | Name | Bemannung Indikator M23042-xx_STA08-xx | |||
| Gewinn | 10.09.2026 | A Schrij | |||
| Gewinn | 25.03.2025 | S Zastraw | |||
| Gewinn | 25.03.2025 | P Mark | |||
| Zeichnungnummer Nr. Nr. Nr. | Produktausgabe Indikator - IND | 2 | |||
| Date | Tagesdatum und Monatdatum Indikator - IND | A3 | |||
2 Metrological characteristics
| Type | MZ3042 |
| Ingres protection rating | IP52 |
| Display | LCD |
| Operating temperature | -10℃/+45℃ |
| Storage temperature | -20℃/+60℃ |
| Accuracy class OIML R76 | III |
| The applied error fraction pi | 0.5 |
| Weighing software version | O.1.yy |
| Application software version | A.x.yy |
| Operating system | Embedded |
| Primary indication unit | kg |
| Weighing range configuration | Single or dual range |
| Max | ≤3000kg |
| Verification Scale interval e | ≥0.5g |
| Actual Scale interval d | ≥0.1g |
| Maximum number of Verification Scale Intervals(per Weighing Range)n | ≤3000 |
| Maximum number of actual scale interval | ≤10000 |
| Maximum subtractive Tare effect | -Max |
Český metrologický institut
-3-
| Minimum signal voltage per verification scale interval(μV/e) | 0,5 |
| Load cell excitation voltage(V) | 3,3 |
| Load cell excitation waveform(V) | DC |
| Minimal load cell impedance(Ω) | 87 |
| Maximal load cell impedance(Ω) | 1200 |
| Load cell connection | 4-wire,6-wire |
| Maximum value of the cable length per cross wire section between the indicator and the junction box or load cells | 50m/mm2In case sense technology is not used the load cells are connected directly without junction box or extension cable |
| High resolution mode | 10x |
| Power supply,voltage | 4x AA 1,5V DC Batteries or6.0VDC,1.0A by AC/DC plug-in power supply or1x Lithium-Ion 3,7V Battery Pack |
3 Devices and functions
determination of stability of equilibrium
indication of stable equilibrium
zero indicating
initial zero setting $ \leq 20% $ Max
zero tracking $ \leq 4 % $ Max
semi-automatic zero setting $ \leq 4 % $ Max
semi-automatic tare balancing (subtractive)
Preset tare device
Hold function
3.1 Additional Instrument Functions
In addition to its primary weighing function, the instrument incorporates supplementary features including:
Body Mass Index (BMI) calculation
Length measurement
Print function
These additional functions have not been tested for compliance.
Český metrologický institut
-3-