R76/2006-A-NL1-19.51
OIML Certificate
OIML Member State The Netherlands
Number R76/2006-A-NL1-19.51 Project number 2383771 Page 1 of 3
Issuing authority NMi Certin B.V. Person responsible: F. van Booma - de Smit
Applicant and Manufacturer Dibal S.A.
Astintze Kalea, 26 - Pol. Ind. Neinver
48160 Derio, Vizcaya
Spain
Identification of the An Indicator certified type Type
Characteristics
This OIML Certificate is issued under scheme A.
See next page
This Certificate attests the conformity of the above identified Type (represented by the sample(s) identified in the OIML Test Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):
OIML R 76 - Edition 2006 for accuracy class or
This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation above-identified. This Certificate does not bestow any form of legal international approval.
Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.
Issuing Authority
NMi Certin B.V., OIML Issuing Authority NL1 22 November 2019
Certification Board
This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.
OIML Certificate
OIML Member State The Netherlands
Number R76/2006-A-NL1-19.51
Project number 2383771
Page 2 of 3
The conformity was established by the results of tests and examinations provided in the associated OIML Type Evaluation Report:
- No. NMi-2382771-01 dated 20 November 2019 that includes 55 pages.
Characteristics of the indicator:
| Accuracy class | or |
| Weighing range(s) | Single interval Multi-interval Multiple range |
| Tare | T≤-Max+e for single interval instruments T≤-Max1+e1 for multi-interval instruments T≤-Max2+e2 for multiple range instruments |
| Maximum number of scale intervals (one weighing range) | n≤6000 |
| Maximum number of scale intervals (multi-interval) | n≤3000 (per partial weighing range) |
| Maximum number of partial weighing ranges | 2 |
| Maximum number of scale intervals (multiple range) | n≤3000 (per weighing range) |
| Maximum number of weighing ranges | 2 |
| Load cell excitation voltage | 5V DC |
| Minimum signal input voltage | Umin=0mV |
| Minimum input voltage per verification scale interval | 1μV |
| Minimum load cell resistance | 21Ω |
| Maximum load cell resistance | 1131Ω |
| Fraction of the maximum permissible error | 0.5 |
| Load cell connection | 6-wire(remote sensing)or4-wire |
| Maximum value of the cable length per cross wire section between the indicator and the junction box or load cells | 2997.9m/mm2 In case a 4-wire connection is used,the load cells are connected directly without junction box |
| Temperature range | 0℃/+40℃ |
| Power supply voltage | 110-230V AC 50/60Hz |
| Application | Can be used for direct sales to the public |
| Firmware identification | 2.0.3 |
OIML Certificate
OIML Member State The Netherlands
Number R76/2006-A-NL1-19.51
Project number 2383771
Page 3 of 3
The indicator has loadable user application software that runs on Linux or Windows system.
PC user software:
| Framework | JPOS | PHP | .NET |
| Software identification | 2.x.x(x=0...9) | 1.x.x(x=0...9) | 3.x.x.x(x=0...9) |
| Checksum | ADD09326 | F244B8F0 | 2ADDCC7C |
The user software identification is displayed permanently on the main screen (for JPOS and PHP) or by pressing "version" on the main screen (for .NET).
The identification number of weighing firmware and user software can both be displayed after opening the application "Test and Adjust" and selecting "Version".