R76/2006-A-NL1-19.23 Rev. 1
OIML Certificate
OIML Member State The Netherlands
Number R76/2006-A-NL1-19.23 Revision 1
Project number 3473231
Page 1 of 3
Issuing authority
NMi Certin B.V. Person responsible: M.Ph.D. Schmidt
Applicant and Manufacturer
Rice Lake Weighing Systems 230 W Coleman St. Rice Lake, WI 54868 United States of America
Identification of the certified type
An Indicator or Terminal Type : 1280 Enterprise series
Characteristics
This OIML Certificate is issued under scheme A.
See next page
This Certificate attests the conformity of the above identified Type (represented by the sample(s) identified in the OIML Test Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):
OIML R 76 - Edition 2006 for accuracy class or
This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation above-identified. This Certificate does not bestow any form of legal international approval.
Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.
Issuing Authority
NMi Certin B.V., OIML Issuing Authority NL1 11 February 2022
Certification Board
This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.
OIML Certificate
OIML Member State The Netherlands
Number R76/2006-A-NL1-19.23 Revision 1
Project number 3473231
Page 2 of 3
The conformity was established by the results of tests and examinations provided in the associated OIML Reports:
No. NMi-14200409-01 dated 29 February 2016 that includes 22 pages;
No. NMi-14200409-02 dated 29 February 2016 that includes 50 pages;
No. NMi-14200409-03 dated 29 February 2016 that includes 22 pages;
No. NMi-2166254-01 dated 10 May 2019 that includes 11 pages.
Characteristics of the indicator:
| Configuration | Analog load cells | Digital load cells or weighing module |
| Accuracy class | III or III | |
| Weighing ranges | Single interval Multi-interval Multiple range | |
| Maximum number of scale intervals (one weighing range) | n≤10000 divisions | - |
| Maximum number of scale intervals (multi-interval) | n≤10000 divisions (per partial weighing range) | - |
| Maximum number of partial weighing ranges | 3 | - |
| Maximum number of scale intervals (multiple range) | n≤10000 divisions (per weighing range) | - |
| Maximum number of weighing ranges | 3 | - |
| Load cell excitation voltage | 10V DC | - |
| Load cell power supply voltage | - | 15V DC |
| Minimum input voltage per verification scale interval | 1,0μV | - |
| Minimum load cell resistance | 23Ω | - |
| Maximum load cell resistance | 1050Ω | - |
| Fraction of the maximum permissible error | 0,5 | 0 |
| Load cell connection | 6 wire with sense technology, may be configured as 4 wire | - |
| Maximum value of the cable length per cross wire section between the indicator and the junction box or load cells | 395m/mm2 In case sense technology is not used the load cells are connected directly without junction box or extension cable | - |
| Maximum number of load platforms | 8 | |
OIML Certificate
OIML Member State The Netherlands
Number R76/2006-A-NL1-19.23 Revision 1
Project number 3473231
Page 3 of 3
| Climatic environment | temperature range | -10℃/+40℃ |
| humidity | non-condensing | |
| intended location | closed | |
| Electromagnetic environment class | E2 | |
| Power supply voltage | 100-240V AC 50/60Hz(forAC version)9-30V DC(forDC version) | |
| Software identification1280 firmware | Version number:V1.xxorV2.xx(xx=00...99for the legally non-relevant software) | |
Software:
The identification number for 1280 firmware will be displayed after pressing the key sequence: Menu -> Audit trail.
The indicator has embedded software.
Revision History
This revision replaces the previous version.
| Revision | Date | Change(s) |
| Initial | 10 May 2019 | - |
| 1 | 11 February 2022 | Terminal option added |