- Applicant
- Mettler-Toledo (Changzhou) Measurement Technology Ltd
- Issuer
- NL1
- Issuing year
- 2023
- Status
- Valid
R76/2006-A-NL1-21.09 Rev. 1
Page 1
O I M L C e r t i f i c a t e
OIML Member State Number R76/2006-A-NL1-21.09 revision 1
The Netherlands Project number 2585293
Page 1 of 3
Issuing authority NMi Certin B.V.
Person responsible: M.Ph.D. Schmidt
Applicant and METTLER-TOLEDO (Changzhou) Measurement Technology Ltd.
Manufacturer No.111, West Taihu Road,
Xinbei district, Changzhou
Jiangsu, 213125
Peoples Republic of China
Identification of the An Indicator
certified type Type : IND231
IND236
IND236-TLF
Characteristics See next page
This OIML Certificate is issued under scheme A.
This Certificate attests the conformity of the above identified Type (represented by the sample(s)
identified in the OIML Test Report) with the requirements of the following Recommendation of the
International Organization of Legal Metrology (OIML):
OIML R 76-1:2006 for accuracy class III or IIII
This Certificate relates only to the metrological and technical characteristics of the type of measuring
instrument covered by the relevant OIML International Recommendation above-identified.
This Certificate does not bestow any form of legal international approval.
Important note: Apart from the mention of the Certificate’s reference number and the name of the
OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of
the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.
Issuing Authority NMi Certin B.V., OIML Issuing Authority NL1
31 October 2023
Certification Board
NMi Certin B.V. This document is issued under the This document is digitally
Thijsseweg 11 provision that no liability is signed and sealed. The
2629 JA Delft accepted and that the applicant digital signature can be
The Netherlands shall indemnify third-party liability. verified in the blue
T +31 88 6362332 ribbon on top of the
certin@nmi.nl The notification of NMi Certin B.V. electronic version of this
www.nmi.nl as Issuing Authority can be verified certificate.
at www.oiml.org
Page 2
O I M L C e r t i f i c a t e
OIML Member State Number R76/2006-A-NL1-21.09 revision 1
The Netherlands Project number 2585293
Page 2 of 3
The conformity was established by the results of tests and examinations provided in the associated
report(s):
- No. NMi-13200101-01 dated 22 April 2013 that includes 49 pages;
- No. NMi-13200101-02 dated 22 April 2013 that includes 13 pages;
- No. NMi-14200515-01 dated 28 November 2014 that includes 12 pages;
- No. NMi-SO15202366-01 dated 26 February 2016 that includes 8 pages.
- No. NMi-2315156-01 dated 8 October 2020 that includes 28 pages.
Characteristics of the indicator:
Accuracy class III or IIII
Single interval
Weighing range(s)
Multiple range
Configuration Without tilt compensation With tilt compensation
Maximum number of scale intervals n 6000 n 1500
Maximum number of weighing ranges 2
Load cell excitation voltage 5 V DC
Minimum signal input voltage Umin = 0 mV Umin = 1 mV
Minimum input voltage per verification
0,5 µV 3,33 µV
scale interval
Minimum load cell resistance 87 175
Maximum load cell resistance 1200
Fraction of the maximum permissible
0,5
error
4-wire or
Load cell connection 6-wire (remote sensing)
6-wire (remote sensing)
No special cable length
Maximum value of the cable length per In case a 4-wire connection
cross wire section between the indicator is used the load cells are 17,3 m/mm2
and the junction box or load cells connected directly without
junction box
Temperature range -10 °C / +40 °C
100 – 240 V AC 50/60 Hz or
internal rechargeable 7,2 V DC battery or
Power supply voltage
9 V DC battery or
12-36 V DC (only for IND236)
Intended to be used for
Application -
on-board weighing
Version number: 1.xx.xxxx
Software identification
(x = 0 … 9)
Page 3
O I M L C e r t i f i c a t e
OIML Member State Number R76/2006-A-NL1-21.09 revision 1
The Netherlands Project number 2585293
Page 3 of 3
Revision History
This revision replaces the previous version.
Revision Date Changes
0 2021-02-08 Initial issue.
1 2023-10-31 Correction for mentioning wrong report number