- Applicant
- Cardinal Scale Manufacturing Co.
- Issuer
- DK2
- Issuing year
- 2026
- Status
- Valid
R76/2006-A-DK2-26.06
Page 1
OIML Member State OIML Certificate No.
Denmark R76/2006-A-DK2-26.06
OIML CERTIFICATE ISSUED UNDER SCHEME A
OIML Issuing Authority
Name: FORCE Certification A/S
Address: Park Allé 345, 2605 Brøndby, Denmark
Person responsible: Per Rafn Crety
Applicant
Name: Cardinal Scale Manufacturing Company
Address: 102 East Daugherty
Webb City, MO 64870
USA
Manufacturer Cardinal Scale Manufacturing Company
Identification of the certified type (the detailed characteristics will be defined in the additional pages)
825
Designation of the module (if applicable)
Non-automatic weighing indicator
This OIML Certificate attests the conformity of the above identified type (represented by the
sample(s) identified in the OIML type evaluation report) with the requirements of the following
Recommendation of the International Organization of Legal Metrology (OIML):
OIML R 76-1, Edition (year): 2006
For accuracy class (if applicable): III
Page 1 of 5
This OIML certificate shall not be reproduced except in full, without written permission by FORCE Certification A/S
FORCE Certification A/S ∙ Park Alle 345 2605 Brøndby Tel+45 43 25 01 77 info@forcecertification.com www.forcecertification.com
Task no.: 124-23866.90.20 and ID no.: FC-OIML-21411-1 forcecertification.com/en/weighing
Page 2
OIML Certificate No.
R76/2006-A-DK2-26.06
This OIML Certificate relates only to metrological and technical characteristics of the type of
measuring instrument covered by the relevant OIML Recommendation identified above.
This OIML Certificate does not bestow any form of legal international approval.
The conformity was established by the results of tests and examinations provided in the associated
OIML reports:
Type Examination Report: No. 124-23866.10 dated 28th of October 2025, that includes 66 pages.
Type Evaluation Report: No. 124-23866.20, dated 02 July 2026, that includes 24 pages.
The technical documentation relating to the identified type is contained in documentation file:
124-23866
OIML Certificate History
Revision No. Date Description of the modification
Initial version 02 of July 2026 -
Identification, signature and stamp
The OIML Issuing Authority
FORCE Certification A/S
Michael Lang Sørensen Date: 02 July 2026
2026-07-02
Digitally signed by Michael Lang Sørensen
mls@forcetechnology.com
Project Manager
Michael Lang Sørensen
Certification Manager
Important note: Apart from the mention of the Certificate’s reference number and the name of the
OIML Member State in which the Certificate is issued, partial quotation of the
Certificate and of the associated OIML type evaluation report(s) is not permitted,
although either may be reproduced in full.
Page 2 of 5
FORCE Certification A/S ∙ Park Alle 345 2605 Brøndby Tel+45 43 25 01 77 info@forcecertification.com www.forcecertification.com
forcecertification.com/en/weighing
Page 3
OIML Certificate No.
R76/2006-A-DK2-26.06
Descriptive annex
Characteristics
Type .................................................. : 825
Accuracy class .................................. : III or IIII
Weighing range ................................ : Single interval or multi-interval (max 3 intervals)
Maximum number of
Verification Scale Intervals .............. : 10,000 (Class III) or 1,000 (Class IIII)
Internal resolution ............................ : >16,000,000 counts
Maximum tare effect ........................ : -Max
Fractional factor ............................... : p’I = 0.5
Minimum input-voltage per VSI ...... : 0.5 µV
Minimum signal voltage for dead
load ................................................... : 1 mV
Excitation voltage ............................. : 10.85 VDC
Analog range .................................... : 1 – 40 mV
Circuit for remote sense ................... : Active
Minimum input impedance .............. : 25 Ω
Maximum input impedance .............. : 1100 ΩMaximum length of lo
Maximum length of load cell cable 46 m/mm2
(6-wire)
Operational temperature ................... : -10° C to +40° C
Mains power supply ......................... : 100 – 240 VAC, 50/60 Hz
Page 3 of 5
FORCE Certification A/S ∙ Park Alle 345 2605 Brøndby Tel+45 43 25 01 77 info@forcecertification.com www.forcecertification.com
forcecertification.com/en/weighing
Page 4
OIML Certificate No.
R76/2006-A-DK2-26.06
Software
The 825 indicator software is segregated into parts, where the legally relevant part is the software that
includes boot loader, standard indicator and streaming of weighing results to RS232, while the
application software apart from display of the weight shall be non-legally relevant in order to be
covered by this type approval certificate.
The software revision level is displayed during the power up sequence of the instrument as
XX.YY.ZZZ where XX is the version of the legally relevant software and YY.ZZZ is the subversion
not affecting the legally functions of the software part.
Software part Approved versions of XX
Boot Loader 1.YY.ZZZ, for YY ≥ 09
Mainboard 1.YY.ZZZ, for YY ≥ 20
OS / Std. Apps 1.YY.ZZZ, for YY ≥ 12
Scale Input Board (SIB) 1.YY.ZZZ, for YY ≥ 07
DIO Board 0.YY.ZZZ, for YY ≥ 05
DAC Board 1.YY.ZZZ, for YY ≥ 01
Devices
• Semi-automatic Zero-setting range: ±2 % of Max.
• Zero-tracking range: ±2 % of Max.
• Initial zero-setting range: ±10 % of Max.
• Semi-automatic subtractive tare
• Preset Tare
• Net/Gross indication
• Printing device (through interface)
• Alibi memory
Page 4 of 5
FORCE Certification A/S ∙ Park Alle 345 2605 Brøndby Tel+45 43 25 01 77 info@forcecertification.com www.forcecertification.com
forcecertification.com/en/weighing
Page 5
OIML Certificate No.
R76/2006-A-DK2-26.06
Interfaces
• Load cell interface
• USB interface
• Logic level inputs
• Analogue output
• Ethernet/LAN
• Wi-Fi
• Bluetooth
• Serial interface
Page 5 of 5
FORCE Certification A/S ∙ Park Alle 345 2605 Brøndby Tel+45 43 25 01 77 info@forcecertification.com www.forcecertification.com
forcecertification.com/en/weighing