- Applicant
- Curiotec Co.,Ltd
- Issuer
- DK2
- Issuing year
- 2024
- Status
- Valid
R76/2006-A-DK2-24.13
Page 1
OIML Member State OIML Certificate No.
Denmark R76/2006-A-DK2-24.13
OIML CERTIFICATE ISSUED UNDER SCHEME A
OIML Issuing Authority
Name: FORCE Certification A/S
Address: Park Allé 345, 2605 Brøndby, Denmark
Person responsible: Per Rafn Crety
Applicant
Name: Curiotec Co. Ltd.,
Address: 79, Myeong-bong-san-ro 352 beon-gil,
guangton-mueon,
Paju-si, Gyeonggi-do, 413-855,
South Korea
Manufacturer Curiotec Co. Ltd.,
CAS (Zhejiang) Electronics Co. Ltd, China.
CAS Corporation, Republic of Korea
CAS Elektronik San. Tic. A.S., Turkey
CAS Deutschland AG, Germany.
Identification of the certified type (the detailed characteristics will be defined in the additional
pages)
CTI-600A and 600D series
Designation of the module (if applicable)
Non-automatic weighing instrument
This OIML Certificate attests the conformity of the above identified type (represented by the
sample(s) identified in the OIML type evaluation report) with the requirements of the following
Recommendation of the International Organization of Legal Metrology (OIML):
OIML R 76-1, Edition (year): 2006
For accuracy class (if applicable): III or IIII
Page 1 of 4 pages
FORCE Certification A/S ∙ Park Alle 345 2605 Brøndby Tel+45 43 25 01 77 info@forcecertification.com www.forcecertification.com
Task no.: 124-31193.90.20 and ID no.: FC-OIML-18526-1 forcecertification.com/en/weighing
Page 2
OIML Certificate No.
R76/2006-A-DK2-24.13
This OIML Certificate relates only to metrological and technical characteristics of the type of
measuring instrument covered by the relevant OIML Recommendation identified above.
This OIML Certificate does not bestow any form of legal international approval.
The conformity was established by the results of tests and examinations provided in the associated
OIML reports:
Type examination report:
TR-661, dated 01 July 2014, that includes 35 pages
No. SN1306, dated 23 March 2015, that includes 10 pages
Type evaluation report: No. 124-31193.90.20, dated 29 November 2024, that includes 17 pages
The technical documentation relating to the identified type is contained in documentation file:
124-31193.10
OIML Certificate History
Revision No. Date Description of the modification
Initial version 12 December 2024 -
Identification, signature and stamp
The OIML Issuing Authority
FORCE Certification A/S
Date: 12 December 2024 Jens Hovgård Jensen
2024-12-12
Digitally signed by Jens Hovgård Jensen
Jens Hovgård Jensen jhje@force.dk
Certification Manager
Certification Manager
Important note: Apart from the mention of the Certificate’s reference number and the name of the
OIML Member State in which the Certificate is issued, partial quotation of the
Certificate and of the associated OIML type evaluation report(s) is not permitted,
although either may be reproduced in full.
Page 2 of 4 pages
Page 3
OIML Certificate No.
R76/2006-A-DK2-24.13
Descriptive annex
Characteristics
The main features of the instruments are,
- Plastic construction
- Keypad with numerical, navigation and function keys
- 4.3 “ LCD display and indicators)
Software
The software is designated “V1.xx”
This information is displayed at power up and may be displayed like this:
where:
- xx is reflecting non-legally relevant changes and may be numbers, letters, symbols or blank,
Access to the legally relevant parameters is only possible by accessing the calibration switch on the
main board. Access to this calibration switch and download of software is prevented by sealing the
enclosure.
Page 3 of 4 pages
Page 4
OIML Certificate No.
R76/2006-A-DK2-24.13
Technical data
Power Supply 100-240VAC 50/60Hz
Maximum number of scale intervals 10,000 for class III
1,000 for class IIII
Maximum Tare value -Max (single interval)
-Maxi (dual interval)
Maximum Preset Tare value -Max (single interval)
-Maxi (dual interval)
Load cell excitation voltage 5 VDC
Minimum load cell impedance 43 Ω
Maximum load cell impedance 1100 Ω
Minimum input voltage per verification scale 0.5 µV
interval
Measuring range minimum voltage 0 mV
Measuring range maximum voltage 16 mV
Fraction of maximum permissible error Pind=0.5
Operating temperature -10°C to -40°C
Load cell cable maximum length 183 m/mm2 (6-wire configuration)
(From indicator to load cell junction box)
Devices
- Initial zero setting device (≤ 20% of Max)
- Semi-automatic zero setting device (≤ 4% of Max)
- Zero tracking device (≤ 4% of Max)
- Zero indicator
- Gross/Net indicator
- Stable indicator
- Semi-automatic subtractive tare balancing device
- Preset Tare
- Gravity compensation
- Hold function
- Soft function keys
- Memory storage
Interfaces
RS232 / RS485.
USB
Relay IN/OUT
Analogue OUT