Skip to content

← R76

R76/2006-A-NL1-18.35

Applicant
Hill-Rom
Issuer
NL1
Issuing year
2018
Status
Valid

Source PDF (certificates/R76/2006/r076-2006-nl1-2018-35.pdf) · extracted by OCR from the digitalized tier of the corpus.

R76/2006-A-NL1-18.35

Page 1

O I M L C e r t i f i c a t e



          OIML Member State                                     Number R76/2006-A-NL1-18.35
           The Netherlands                                                        Project number 1902117
                                                                      Page 1 of 2


              Issuing authority      NMi Certin B.V.
                                  Person responsible: C. Oosterman

             Applicant and          Hill-Rom, Inc.
            Manufacturer         1069 State Route 46 East
                                          Batesville, IN 47006
                                  United States of America

              Identification of the   A Non-automatic weighing instrument
               certified type         Type                                                        :  Centrella

               Characteristics         See next page


              This OIML Certificate is issued under scheme A.

              This Certificate attests the conformity of the above identified Type (represented by the sample(s)
              identified in the OIML Test Report) with the requirements of the following Recommendation of the
              International Organization of Legal Metrology (OIML):

                           OIML R 76 - Edition 2006 for accuracy class  III  or  IIII

              This Certificate relates only to the metrological and technical characteristics of the type of measuring
             instrument covered by the relevant OIML International Recommendation above-identified.
              This Certificate does not bestow any form of legal international approval.

            Important note: Apart from the mention of the Certificate’s reference number and the name of the
          OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of
            the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.





              Issuing Authority     NMi Certin B.V., OIML Issuing Authority NL1
                                16 October 2018




                                        C. Oosterman
                             Head Certification Board

              NMi Certin B.V.                    This document is issued under the
              Hugo de Grootplein 1             provision that no liability is
                 3314 EG Dordrecht              accepted and that the applicant
                  the Netherlands                    shall indemnify third-party liability.
                 T +31 78 6332332
                   certin@nmi.nl                 The notification of NMi Certin B.V.
                www.nmi.nl                      as Issuing Authority can be verified
                                                      at www.oiml.org


101

Page 2

O I M L C e r t i f i c a t e


          OIML Member State                                     Number R76/2006-A-NL1-18.35
           The Netherlands                                                        Project number 1902117
                                                                      Page 2 of 2


           The conformity was established by the results of tests and examinations provided in the associated
          OIML Type Evaluation Report(s):

                   -   No. NMi-1902117-01 dated 16 October 2018 that includes 50 pages.

             Characteristics of the non-automatic weighing instrument:

             Accuracy class                                                                           III  or  IIII
          Maximum capacity                                Max  250 kg
               Verification scale interval                                   e  0,5 kg
            Weighing range(s)                                             Single interval
          Maximum number of scale intervals
                                                            n  500 divisions             (one weighing range)
            Temperature range                                     +5 °C / +35 °C
           Power supply voltage                                100 - 240 V AC 50/60 Hz
              Application                                  Intended to be used for weighing patients
             Software identification                                    V2





101