- Applicant
- Rice Lake Weighing Systems
- Issuer
- NL1
- Issuing year
- 2018
- Status
- Valid
R76/2006-A-NL1-18.23
Page 1
O I M L C e r t i f i c a t e
OIML Member State Number R76/2006-A-NL1-18.23
The Netherlands Project number 1902470
Page 1 of 2
Issuing authority NMi Certin B.V.
Person responsible: C. Oosterman
Applicant and Rice Lake Weighing Systems
Manufacturer 230 West Coleman Street
Rice Lake, WI54868
United States of America
Identification of the An Indicator
certified type Type : 880-2A, 880-2D, 880-3A, 880-3D
Characteristics See next page
This OIML Certificate is issued under scheme A.
This Certificate attests the conformity of the above identified Type (represented by the sample(s)
identified in the OIML Test Report) with the requirements of the following Recommendation of the
International Organization of Legal Metrology (OIML):
OIML R 76 - Edition 2006 for accuracy class III or IIII
This Certificate relates only to the metrological and technical characteristics of the type of measuring
instrument covered by the relevant OIML International Recommendation above-identified.
This Certificate does not bestow any form of legal international approval.
Important note: Apart from the mention of the Certificate’s reference number and the name of the
OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of
the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.
Issuing Authority NMi Certin B.V., OIML Issuing Authority NL1
17 May 2018
C. Oosterman
Head Certification Board
NMi Certin B.V. This document is issued under the
Hugo de Grootplein 1 provision that no liability is
3314 EG Dordrecht accepted and that the applicant
the Netherlands shall indemnify third-party liability.
T +31 78 6332332
certin@nmi.nl The notification of NMi Certin B.V.
www.nmi.nl as Issuing Authority can be verified
at www.oiml.org
101
Page 2
O I M L C e r t i f i c a t e
OIML Member State Number R76/2006-A-NL1-18.23
The Netherlands Project number 1902470
Page 2 of 2
The conformity was established by the results of tests and examinations provided in the associated
OIML Test Reports:
- No. NMi-13200624-01 dated 13 June 2014 that includes 50 pages;
- No. NMi-13200624-02 dated 26 June 2014 that includes 8 pages.
Characteristics of the indicator:
Configuration Analog load cells
Accuracy class OIML R 76 III or IIII
Single interval
Weighing range(s) Multi-interval
Multiple range
Maximum number of scale intervals
n 10000 divisions (one weighing range)
Maximum number of scale intervals n 10000 divisions
(multi-interval) (per partial weighing range)
Maximum number of
2
partial weighing ranges
Maximum number of scale intervals n 10000 divisions
(multiple range) (per weighing range)
Maximum number of weighing ranges 2
Load cell excitation voltage 10 V DC
Minimum input voltage Umin = 0 mV
Minimum input voltage per verification
1 µV
scale interval
Minimum load cell resistance 21
Maximum load cell resistance 1050
Fraction of the maximum permissible error 0,5
Load cell connection 6-wire (remote sensing)
Maximum value of the cable length per No special cable length
cross wire section between the indicator and In case a 4-wire connection is used the load cells are
the junction box or load cells connected directly without junction box
Temperature range -10 °C / +40 °C
100 – 240 V AC 50/60 Hz
Power supply voltage 9 – 36 V DC
(not suitable for a road vehicle power supply)
Version number: 1.xx
Software identification (xx is a number between 00 and 99)
(1. represents the legally relevant software)
101