- Applicant
- Flintec UK Limited
- Issuer
- DK2
- Issuing year
- 2022
- Status
- Valid
R76/2006-A-DK2-22.03
Page 1
OIML Member State OIML Certificate No.
Denmark R76/2006-A-DK2-2022.03
OIML CERTIFICATE ISSUED UNDER SCHEME A
OIML Issuing Authority
Name: FORCE Certification A/S
Address: Park Allé 345, 2605 Brøndby, Denmark
Person responsible: Per Rafn Crety
Applicant
Name: Flintec UK Ltd.
Address: Caxton House,
Caxton Place,
Pentwyn
Cardiff CF23 8HG,
United Kingdom
Manufacturer Flintec UK Ltd.
Identification of the certified type (the detailed characteristics will be defined in the additional
pages)
ER500-G
Designation of the module (if applicable)
Non-automatic weighing indicator / transmitter
This OIML Certificate attests the conformity of the above identified type (represented by the
sample(s) identified in the OIML type evaluation report) with the requirements of the following
Recommendation of the International Organization of Legal Metrology (OIML):
OIML R 76-1, Edition (year): 2006
For accuracy class (if applicable): III and IIII
Page 1 of 6 pages
FORCE Certification A/S ∙ Park Alle 345 2605 Brøndby Tel+45 43 25 01 77 info@forcecertification.com www.forcecertification.com
Task no.: 121-24186.90.50 and ID no.: FC-OIML-12110-1 forcecertification.com/en/weighing
Page 2
OIML Certificate No.
R76/2006-A-DK2-2022.03
This OIML Certificate relates only to metrological and technical characteristics of the type of
measuring instrument covered by the relevant OIML Recommendation identified above.
This OIML Certificate does not bestow any form of legal international approval.
The conformity was established by the results of tests and examinations provided in the associated
OIML reports:
Type examination report: No. 121-24186.10, dated 06 August 2021, that includes 66 pages
Type evaluation report: No. 121-24186.90.50, dated 09 May 2022, that includes 17 pages
The technical documentation relating to the identified type is contained in documentation file:
121-24186
OIML Certificate History
Revision No. Date Description of the modification
Initial version 16 June 2022 -
Identification, signature and stamp
The OIML Issuing Authority
FORCE Certification A/S
Date: 16 June 2022 Jens Hovgård Jensen 2022-06-16
Digitally signed by Jens Hovgård Jensen
jhje@force.dk
Certification Manager
Jens Hovgård Jensen
Certification Manager
Important note: Apart from the mention of the Certificate’s reference number and the name of the
OIML Member State in which the Certificate is issued, partial quotation of the
Certificate and of the associated OIML type evaluation report(s) is not permitted,
although either may be reproduced in full.
Page 2 of 6 pages
Page 3
OIML Certificate No.
R76/2006-A-DK2-2022.03
Descriptive annex
Characteristics
Type: ER500-G
Accuracy class: III and IIII
Weighing range: Single-interval, multi-range or multi-interval
Maximum number of Verification Sca1e Intervals: 10000 (class III), 1000 (class IIII) or
3×10000 (class III), 3×1000 (class IIII)
Maximum tare effect: -Max for single-interval and multi-range
-Max1 for multi-interval
Fractional factor: p'i = 0.5
Minimum input-voltage per VSI: 0.2 µV
Minimum signal voltage for dead load: 0 mV
Excitation voltage: 5 VDC
Load cell interface: 4-wire or 6-wire
Minimum input-impedance: 43 ohm
Maximum input-impedance: 1200 ohm
Mains power supply: 9-32 VDC - not to be supplied from DC mains.
Operational temperature: -15 °C to +55 °C
Maximum cable length between ER500 and
junction box 1533 m/mm2
Software
The software version is displayed during the start-up of the indicator. (Alternating with the TAC
number). The version format is xx.yy.zz, where x is the basic software family, while yy is version
numbers for minor legally relevant changes and zz is changes and corrections not influencing the legal
function of the software.
The approved software version is 01.01.zz.
TAC number
The non-resettable Traceable Access Code is displayed during the start-up of the indicator in the
format: xxxxx. (Alternating with the software version).
Devices
- Initial zero setting device (≤ 20% of Max)
- Semi-automatic zero setting device (≤ 4% of Max)
- Zero tracking device (≤ 4% of Max)
- Semi-automatic subtractive tare balancing device
- Units (Allowed units are g, kg and t.)
- Stable equilibrium, Zero and Net indicators.
Page 3 of 6 pages
Page 4
OIML Certificate No.
R76/2006-A-DK2-2022.03
Interfaces
- RS232
RS-485 / RS-422
USB
Ethernet
3 logical inputs
3 logical outputs
Analogue Output
The interfaces do not have to be secured.
Sealing
Access to the set-up and calibration facility requires that a calibration jumper is removed from the
main board. The jumper can be accessed from the outside, top part of the housing.
The indicator also has a Traceable Access Counter, which increment each time the calibration or legal
part of the set-up has been changed.
The sealing of the calibration jumper, which also prevents the housing from being dismantled - is
accomplished with a brittle plastic sticker. The sticker is placed across the opening designated 30
behind which the calibration jumper is located.
Page 4 of 6 pages
Page 5
OIML Certificate No.
R76/2006-A-DK2-2022.03
Figure 1 Dimensions of ER500-G.
Figure 2 Front panel display with LED indicators and buttoms.
Page 5 of 6 pages
Page 6
OIML Certificate No.
R76/2006-A-DK2-2022.03
Figure 3 Top side label of ER500-G.