R61/2017-A-NL1-23.02 Rev. 0

OIML Certificate

OIML Member State The Netherlands

Number R61/2017-A-NL1-23.02 revision 0

Project number 3469362

Page 1 of 4

Issuing authority

NMi Certin B.V. Person responsible: M.Ph.D. Schmidt

Applicant and Manufacturer

Yamato Scale Co., Ltd.

5-22, Saenba-cho

Akashi, 673-8688

Japan

Identification of the certified type

An Automatic gravimetric filling instrument Type : TSD-N3

Characteristics

This OIML Certificate is issued under scheme A.

See next page

This Certificate attests the conformity of the above identified Type (represented by the sample(s) identified in the OIML Test Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

OIML R 61-1:2017 for reference class Ref (0,5)

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation above-identified. This Certificate does not bestow any form of legal international approval.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 13 June 2023

Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

OIML Certificate

OIML Member State The Netherlands

Number R61/2017-A-NL1-23.02 revision 0

Project number 3469362

Page 2 of 4

The conformity was established by the results of tests and examinations provided in the associated OIML Type Evaluation Reports:

No. NMi-3469362-01 dated 2 June 2023 that includes 35 pages;

No. NMi-3469362-02 dated 2 June 2023 that includes 22 pages;

No. NMi-3469362-03 dated 2 June 2023 that includes 19 pages.

Characteristics of the automatic gravimetric filling instrument

Method of operationselective combination weighing
Reference accuracy classRef(0,5) the operational accuracy classX(x)是 determined at the time of putting into use
Electromagnetic environment classE2
Climatic environmenttemperature range0℃/+40℃
humiditynon-condensing
intended locationClosed
Maximum capacity (of each load receptor)Max≤1.5kg
Minimum capacity (of each load receptor)Min≥8g
Rated minimum fillMinFill≥Min
MaximumfillMaxfill是 determined at the time of putting into use
Maximum number of load receptors16
Power supply voltage200-240V AC 50/60Hz
Software identification A/D board(ADPD)Version:1.01.00 Checksum:A71E0E37
Software identification Main board(ADPD)Version:1.02.00 Checksum:8CA93123
Software identification DDPDVersion:1.02.00 Checksum:4C2C3174

Software:

  • The identification number will be displayed after pressing the key sequence:

  • From the main menu choose "trouble shooting";

  • Select "software version".

OIML Certificate

OIML Member State The Netherlands

Number R61/2017-A-NL1-23.02 revision 0

Project number 3469362

Rated minimum fill (Minfill):

Page 3 of 4

Average number of fills=2Average number of fills=3
d[g]X(0,5)X(1)X(2)X(0,5)X(1)X(2)
131168381910
212632161543820
547015540115519050
10189047016023101150190
Average number of fills=4Average number of fills=5
d[g]X(0,5)X(1)X(2)X(0,5)X(1)X(2)
1442211502512
217844221985024
513353351101490375125
102670133033029801490370
Average number of fills=6Average number of fills=7
d[g]X(0,5)X(1)X(2)X(0,5)X(1)X(2)
110927141182915
23261082835211830
516354101351765440145
103270163041035301760440

OIML Certificate

OIML Member State The Netherlands

Number R61/2017-A-NL1-23.02 revision 0 Project number 3469362 Page 4 of 4

Page 4 of 4

Revision History

RevisionDateChange(s)