R61/2017-A-NL1-22.01 Rev. 0
OIML Certificate
OIML Member State The Netherlands
Number R61/2017-A-NL1-22.01 revision 0 Project number 3269935 Page 1 of 4
Issuing authority
NMi Certin B.V. Person responsible: M.Ph.D. Schmidt
Applicant and Manufacturer
PIM Products B.V.
Herenweg 16
8536 TN Oosterzee
The Netherlands
Identification of the certified type
An Automatic gravimetric filling instrument
Type
Characteristics
: PIM MHWx (where x represents the number of load receptors)
See next page
This OIML Certificate is issued under scheme A.
This Certificate attests the conformity of the above identified Type (represented by the sample(s) identified in the OIML Test Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):
OIML R 61-1:2017 for reference class Ref (1)
This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation above-identified. This Certificate does not bestow any form of legal international approval.
Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.
Issuing Authority
NMi Certin B.V., OIML Issuing Authority NL1 21 December 2022
Certification Board
This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.
OIML Certificate
OIML Member State The Netherlands
Number R61/2017-A-NL1-22.01 revision 0 Project number 3269935 Page 2 of 4
The conformity was established by the results of tests and examinations provided in the associated OIML Type Evaluation Reports:
No. NMi-3269935-01 dated 21 December 2022 that includes 31 pages;
No. NMi-3269935-02 dated 21 December 2022 that includes 19 pages;
No. NMi-3269935-03 dated 21 December 2022 that includes 28 pages.
Characteristics of the automatic gravimetric filling instrument
| Method of operation | selective combination weighing or selective combination weighing combined with cummulative filling or filling by one weighing cycle | ||
| Reference accuracy class | Ref(1) the operational accuracy class X(x) is determined at the time of putting into use | ||
| Electromagnetic environment class | E2 | ||
| Climatic environment | temperature range | -10℃/+40℃ | |
| humidity | ADPD and DDPD | Mechanical assembly | |
| non-condensing | condensing | ||
| intended location | Closed | open and closed | |
| Maximum capacity of each load receptor | Calculated using the compatibility of modules form, contained in OIML R76(2006) clause F.4 | ||
| Scale interval | Calculated using the compatibility of modules form, contained in OIML R76(2006) clause F.4 | ||
| Number of load receptors | 4 to 24 | ||
| Power supply voltage | 24V DC | ||
| Software identification Monitor SW-version | V7,xxx(xxx is number between 001 and 999) | ||
| Software identification BrainsBox SW-versie | zzzz-V7.xxx:yyyy-mm-dd(zzzz is number between 0000 and 9999)(xxx is number between 013 and 999)yyyy-mm-dd represents a date | ||
The identification number will be displayed on the Human Machine Interface (HMI) by:
Selecting the button in the lower left corner of the main menu;
Selecting the information button.
OIML Certificate
OIML Member State The Netherlands
Number R61/2017-A-NL1-22.01 revision 0 Project number 3269935
Page 3 of 4
Characteristics of the Analog Data Processing Device
| Accuracy class | OIML R 76 | or |
| OIML R 61 | Ref(0,2) | |
| Weighing range | Single interval | |
| Maximum number of scale intervals | n≤3000 | |
| Load cell excitation voltage | 5V AC square wave | |
| Minimum signal input voltage | Umin=0mV | |
| Minimum input voltage per verification scale interval | 1μV | |
| Minimum load cell resistance | 350Ω | |
| Maximum load cell resistance | 1050Ω | |
| Fraction of the maximum permissible error | 0,5 | |
| Load cell interface | 6-wire with sense technology, may be configured as 4-wire | |
| Maximum value of the cable length per cross wire section between the indicator and the junction box or load cells | No special cable lengthIn case sense technology is not used the load cells are connected directly without junction box or extension cable | |
Rated minimum fill (Minfill):
| Minfill [g] | ||||||||
| Operational accuracy class X(1) | ||||||||
| Average number of loads per fill | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
| 0.5g | 22 | 31.5 | 38.5 | 44.5 | 49.5 | 109 | 117.5 | 125.5 |
| 1g | 44 | 126 | 154 | 178 | 199 | 327 | 353 | 377 |
| 2g | 178 | 378 | 462 | 1066 | 1192 | 1306 | 1412 | 1508 |
| 5g | 1335 | 1885 | 2310 | 2665 | 2980 | 3265 | 3530 | 3770 |
| 10g | 2670 | 3770 | 4620 | 5330 | 5960 | 6530 | 7060 | 7540 |
OIML Certificate
OIML Member State The Netherlands
Number R61/2017-A-NL1-22.01 revision 0 Project number 3269935 Page 4 of 4
Page 4 of 4
| Minfill [g] | ||||||||
| Operational accuracy class X(2) | ||||||||
| Average number of loads per fill | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
| 0.5g | 11 | 15.5 | 19 | 22 | 25 | 27 | 29.5 | 31.5 |
| 1g | 22 | 31 | 38 | 44 | 50 | 109 | 118 | 126 |
| 2g | 44 | 126 | 154 | 178 | 198 | 326 | 352 | 378 |
| 5g | 335 | 470 | 1155 | 1335 | 1490 | 1635 | 1765 | 1885 |
| 10g | 1330 | 1890 | 2310 | 2670 | 2980 | 3270 | 3530 | 3770 |
Revision History
| Revision | Date | Change(s) |
| 0 | 2022-12-21 | Initial issue. |