R61/2017-A-DK2-25.01

OIML Certificate No. R61/2017-A-DK2-25.01

OIML CERTIFICATE ISSUED UNDER SCHEME A

OIML CERTIFICATE ISSUED UNDER SCHEME A
OIML Issuing Authority
Name: FORCE Certification A/S Address: Park Allé 345, 2605 Brøndby, Denmark Person responsible: Per Rafn Crety
Applicant
Name: Flintec UK Ltd. Address: Caxton House Caxton Place, Pentwyn, Cardiff CF23 8HG United Kingdom
Manufacturer: Flintec Transducers Pvt Ltd. Katunayake, Sri Lanka
Identification of the certified type (the detailed characteristics will be defined in the additional pages) EM100-F
Designation of the module (if applicable) Automatic Gravimetric filling instrument
This OIML Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML type evaluation report) with the requirements of the following Recommendation of the International Organization of Legal Metrology(OIML): OIML R 61-1, Edition(year):2017 For accuracy class(if applicable):Ref(0.2)
OIML Certificate No. R61/2017-A-DK2-25.01
This OIML Certificate relates only to metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML Recommendation identified above.
This OIML Certificate does not bestow any form of legal international approval.
The conformity was established by the results of tests and examinations provided in the associated OIML reports:
Type examination report: No. DANAK-1918640, dated 6 December 2017, that includes 76 pages.
Type examination report: No. DANAK-1918824, dated 26 January 2018, that includes 60 pages
Type evaluation report: No. 124-34253.90.70, dated 13 May 2025, that includes 24 pages,
The technical documentation relating to the identified type is contained in documentation file:124-34253
OIML Certificate History
Revision No.DateDescription of the modification
Initial version15 May 2025-
Identification, signature and stampThe OIML Issuing AuthorityFORCE Certification A/S
Date:15 May 2025Jens Hovgård Jensen2025-05-15Digitally signed by Jens Hovgård Jensenjhje@force.dkCertification Manager
Important note: Apart from the mention of the Certificate's reference number and the name of theOIML Member State in which the Certificate is issued, partial quotation of theCertificate and of the associated OIML type evaluation report(s) is not permitted,although either may be reproduced in full.

Descriptive annex

Characteristics

Type: EM100-F load cell digitizing unit.

Reference class: Ref(0.2)

Accuracy class: 0.2 or 0.5 or 1 or 2

Weighing range: Single-interval, multi-range or multi-interval

Maximum number of verification scale intervals (n): 10,000

Minimum input voltage per VSI ( $ \mathrm{e_{i}} $ ): 0.3 $ \mu\mathrm{V} $

Maximum capacity of interval (Max $ \mathrm{i} $ ): $ n_{\mathrm{i}} \times e_{\mathrm{i}} $

Initial zero-setting range: 20 % of Max

Maximum tare effect: 100 % of Max

Fractional factor ( $ p_{\mathrm{i}} $ ): 0.5

Excitation voltage: 5 VDC

Minimum input voltage from load cell: 0 mV

Maximum input voltage from load cell: 15 mV

Circuit for remote sense: Active (see below)

Minimum input impedance: 58 Ohm

Maximum input impedance: 1100 Ohm

Load cell linearization feature: None

Connecting cable to load cell(s): See Section 3.1.1

Maximum time between automatic zero-setting: 100 minutes $ ^{1)} $

Minimum warm-up time: 15 minutes $ ^{1)} $

Supply voltage: 9-32 VDC, not to be supplied from DC Mains

Operating temperature range:

Peripheral interface(s):

MinFill's dependency of accuracy class and d in g for $ \mathrm{d}=0. 3 \ \mu\mathrm{V}. $ $ ^{1)} $

d[g]Accuracy class
X(0.2)X(0.5)X(1)X(2)
d[kg]d[kg]d[kg]d[kg]
0.1560.0056220.0022110.001160.0006
0.2560.0112220.0044110.002260.0012
0.5560.028220.011110.005560.0030
11110.111220.022110.01160.006
21670.334220.044110.02260.012
53331.665670.335220.1160.030
103333.331331.33330.33110.110
203336.661332.66671.34170.340
50500251336.65673.35331.65
1005005020020676.7333.3
2005001002004010020336.6
≥50050020010050

$ ^{1)} $ Values applies for single fill only.

Software

The software version may be viewed by sending "FFV" to the unit, which responds with 'Vxx.yy' or 'V:xx.yy.zz'.

where xx denotes the legally relevant code, yy denotes the major non-legally relevant code, and zz denotes the minor non-legally relevant code.

The tested software version is: 'V01.01'.

The software changes from V01.01 to V02.00.00 have been examined.

Devices

  • Initial zero-setting

  • Semi-automatic zero-setting

  • Zero tracking

  • Semi-automatic subtractive tare

  • Preset tare

  • Automatic tare

  • Event counter (TAC)

Peripheral interfaces

  • RS485

  • RS232

  • CANopen

  • 2 logic level inputs

  • 2 open-drain outputs

The peripheral interfaces are characterised "Protective interfaces" according to paragraph 8.4 in the Directive.