R61/2017-A-DK2-22.01 Rev. 1

OIML Certificate No. R61/2017-A-DK2-22.01 Revision 1

OIML CERTIFICATE ISSUED UNDER SCHEME A

OIML CERTIFICATE ISSUED UNDER SCHEME A
OIML Issuing Authority
Name: FORCE Certification A/S Address: Park Allé 345, 2605 Brøndby, Denmark Person responsible: Per Rafn Crety
Applicant
Name: Flintec UK Ltd. Address: Caxton House, Caxton Place, Pentwyn Cardiff CF23 8HG, United Kingdom
Manufacturer: Flintec UK Ltd.
Identification of the certified type (the detailed characteristics will be defined in the additional pages) ER500-F
Designation of the module (if applicable) Weighing transmitter for automatic gravimetric filling instruments
This OIML Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML type evaluation report) with the requirements of the following Recommendation of the International Organization of Legal Metrology(OIML): OIML R 61-1, Edition(year):2017 For accuracy class(if applicable):Ref(0.5)
Revision No.DateDescription of the modification
Initial version17 June 2022-
Revision 124 October 2024New software version added.

Descriptive annex

Characteristics
Type:ER500-F
Reference class:Ref(0.5)
Accuracy class:0.5,1 or 2
Single-interval, multi-range or multi-interval
Maximum number of verification scale intervals(ni):≤3×10000
Maximum tare effect:-Max
Fractional factor:p'i=0.5
Minimum input-voltage per VSI:0.2μV
MinFill:See tables below
Extra warm-up time:3 minutes
Maximum time between automatic zero-setting:120 minutes
Excitation voltage:5VDC
Load cell interface:4-wire or 6-wire
Minimum input-impedance:43ohm
Maximum input-impedance:1200ohm
Mains power supply:9-32VDC- not to be supplied from DC mains.
Operational temperature:-15℃ to +55℃
Maximum cable length between ER500-F and junction box1533m/mm2

MinFill for 0.2 $ \mu\mathrm{V} \leq\mathrm{d} < 0. 4 \ \mu\mathrm{V} $

dX(0.5)X(1)X(2)
[g]d[g]d[g]d[g]
0,128128,1707181,8
0,2561112,27014183,6
0,58424217035189
1168316831401401818
2168333662104201836
516838415420210035175
1025252525042042001051050
2025255050042084001052100
502525126250630315001055250
10025252525006306300015815800
200252550500063012600015831600
5002525126250063031500015879000

MinFill for 0.4 $ \mu\mathrm{V} \leq\mathrm{d} < 0. 8 \ \mu\mathrm{V} $

dX(0.5)X(1)X(2)
[g]d[g]d[g]d[g]
0,114114,1353,590,9
0,214128,235791,8
0,5281140,53517,594,5
1421421353599
2842168470140918
584242102101050945
108428420210210018180
201263252602104200531060
5012636315031515750532650
100126312630031531500535300
2001263252600315630007915800
50012636315003151575007939500

MinFill for 0.8 $ \mu\mathrm{V} \leq\mathrm{d} $

dX(0.5)X(1)X(2)
[g]d[g]d[g]d[g]
0,1717,1181,860,6
0,27114,2183,661,2
0,57135,518963
1141141181866
22114221836612
5421210535175630
104214210105105011110
204218420105210017340
50632316001055250331650
1006326320015815800333300
20063212640015831600336600
500632316000158790005025000

Software

The software version is displayed during the start-up of the indicator. (Alternating with the TAC number). The version format is xx.yy.zz, where xx denotes the version of the legally relevant code, while yy is the version numbers for major non-legally relevant changes and zz is the version numbers for minor non-legally relevant changes.

The approved software versions are:

Main firmware: 01.yy.zz

Main firmware: 02.yy.zz

Display firmware: 02.yy.zz

TAC number

The non-resettable Traceable Access Code is displayed during the start-up of the indicator in the format: xxxxx. (Alternating with the software version).

Devices

  • Initial zero setting device ( $ \leq 20% $ of Max)

  • Semi-automatic zero setting device ( $ \leq 4 % $ of Max)

  • Zero tracking device ( $ \leq 4 % $ of Max)

  • Semi-automatic subtractive tare balancing device

  • Units (Allowed units are g, kg and t.)

  • Stable equilibrium, Zero and Net indicators.

Display and buttoms

If the ER500-F is equipped with a display and buttoms these are are for set-up and service purpose only.

Interfaces

  • RS232

RS-485/RS-422

USB

Ethernet

3 logical inputs

3 logical outputs

Analogue Output

The interfaces do not have to be secured.

Sealing

Access to the set-up and calibration facility requires that a calibration jumper is removed from the main board. The jumper can be accessed from the outside, top part of the housing.

The indicator also has a Traceable Access Counter, which increment each time the calibration or legal part of the set-up has been changed.

The sealing of the calibration jumper, which also prevents the housing from being dismantled - is accomplished with a brittle plastic sticker. The sticker is placed across the opening designated 30 behind which the calibration jumper is located.

Figure 2 Front panel display with LED indicators and buttoms.

Figure 3 Top side label of ER500-F.