Skip to content

← R60

R60/2021-A-NL1-26.09 Rev. 1

Applicant
Mettler-Toledo AG
Issuer
NL1
Issuing year
2026
Status
Valid

Source PDF (certificates/R60/2021/r60-2021-nl1-2026-09-rev1.pdf) · extracted by OCR from the digitalized tier of the corpus.

R60/2021-A-NL1-26.09 Rev. 1

OIML Certificate

OIML Member State The Netherlands

Number R60/2021-A-NL1-26.09 revision 1 Project number 3940755 Page 1 of 3

Issuing authority NMi Certin B.V. Person responsible: M.Ph.D. Schmidt

Applicant Mettler-Toledo GmbH Im Langacher 44 8606 Greifensee Switzerland

Manufacturer

Mettler-Toledo (Changzhou) Precision Instruments Ltd No.22, Zhengqiang Road, Xinbei District Changzhou, Jiangsu P.R. China

Identification of the certified type A single point load cell, with strain gauges, equipped with electronics. Registered trade name : METTLER TOLEDO

Type

: SLP86xD (x=1,2,3)

Characteristics

See next page

This OIML Certificate is issued under scheme A.

This Certificate attests the conformity of the above identified Type (represented by the sample(s) identified in the OIML Test Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

OIML R 60-1:2021 for accuracy class C

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation above-identified. This Certificate does not bestow any form of legal international approval.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 29 June 2026

Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

Reproduction of the complete document only is permitted.

OIML Certificate

OIML Member State The Netherlands

Number R60/2021-A-NL1-26.09 revision 1

Project number 3940755

Page 2 of 3

The conformity was established by the results of tests and examinations provided in the associated reports:

No. NMi-3940755-01 dated 15 June 2026 that includes 71 pages;

No. NMi-3940755-02 dated 15 June 2026 that includes 46 pages;

  • No. NMi-3940755-03 dated 15 June 2026 that includes 46 pages;

  • No. NMi-3940755-04 revision 1 dated 29 June 2026 that includes 25 pages.

Characteristics of the load cell:

Characterization of load cell capabilitiesDigital load cell with data processing
Maximum capacity ($E_{\mathrm{max}}$)10kg up to and including 750kg
Minimum dead load0kg
Accuracy ClassC
Maximum number of load cell intervals(n)^{(1)}$10000
Ratio of minimum LC Verification interval$^{(1)}$ $Y=E_{\mathrm{max}}/v_{\mathrm{min}}$40000
Ratio of minimum dead load output return$^{(1)}$ $Z=E_{\mathrm{max}}/(2*\mathrm{DR})$10000
Temperature range-10℃/+40℃
Fraction $p_{\mathrm{LC}}$0,8
Humidity ClassCH
Safe overload150% of $E_{\mathrm{max}}$
Recommended excitation12-24V DC
Excitation maximum30V DC
Transducer materialStainless steel
Atmospheric protectionHermetically welded
Electromagnetic environment classE2
Number of counts for $E_{\mathrm{max}}$≥Y*5/pLC
Software identificationDescribed in the certificate TC8039

Remark:

  1. The characteristics for $ n_{\mathrm{max}} $ , Y and Z can be reduced separately.

Each load cell produced is provided with an accompanying document with information about its characteristics.

OIML Certificate

OIML Member State The Netherlands

Number R60/2021-A-NL1-26.09 revision 1

Project number 3940755

Page 3 of 3

The above identified Type (represented by the sample(s) identified in the OIML Test Report) have been found to comply with the additional national requirements established by the United States of America (NIST Handbook 44 and NCWM Publication 14), included in the Utilizer Declaration:

  • R 60 OIML-CS rev.2 Additional requirements from the United States Accuracy class III L;

  • R 60 OIML-CS rev.2 Additional requirements from the United States Marking requirements

Revision History

RevisionDateChange(s)
02026-06-15Initial issue.
12026-06-29Test report NMi-3940755-04 replaced due to correction in it.