R60/2021-A-NL1-26.09 Rev. 1
OIML Certificate
OIML Member State The Netherlands
Number R60/2021-A-NL1-26.09 revision 1 Project number 3940755 Page 1 of 3
Issuing authority NMi Certin B.V. Person responsible: M.Ph.D. Schmidt
Applicant Mettler-Toledo GmbH Im Langacher 44 8606 Greifensee Switzerland
Manufacturer
Mettler-Toledo (Changzhou) Precision Instruments Ltd No.22, Zhengqiang Road, Xinbei District Changzhou, Jiangsu P.R. China
Identification of the certified type A single point load cell, with strain gauges, equipped with electronics. Registered trade name : METTLER TOLEDO
Type
: SLP86xD (x=1,2,3)
Characteristics
See next page
This OIML Certificate is issued under scheme A.
This Certificate attests the conformity of the above identified Type (represented by the sample(s) identified in the OIML Test Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):
OIML R 60-1:2021 for accuracy class C
This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation above-identified. This Certificate does not bestow any form of legal international approval.
Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.
Issuing Authority
NMi Certin B.V., OIML Issuing Authority NL1 29 June 2026
Certification Board
This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.
Reproduction of the complete document only is permitted.
OIML Certificate
OIML Member State The Netherlands
Number R60/2021-A-NL1-26.09 revision 1
Project number 3940755
Page 2 of 3
The conformity was established by the results of tests and examinations provided in the associated reports:
No. NMi-3940755-01 dated 15 June 2026 that includes 71 pages;
No. NMi-3940755-02 dated 15 June 2026 that includes 46 pages;
No. NMi-3940755-03 dated 15 June 2026 that includes 46 pages;
No. NMi-3940755-04 revision 1 dated 29 June 2026 that includes 25 pages.
Characteristics of the load cell:
| Characterization of load cell capabilities | Digital load cell with data processing |
| Maximum capacity ($E_{\mathrm{max}}$) | 10kg up to and including 750kg |
| Minimum dead load | 0kg |
| Accuracy Class | C |
| Maximum number of load cell intervals(n)^{(1)}$ | 10000 |
| Ratio of minimum LC Verification interval$^{(1)}$ $Y=E_{\mathrm{max}}/v_{\mathrm{min}}$ | 40000 |
| Ratio of minimum dead load output return$^{(1)}$ $Z=E_{\mathrm{max}}/(2*\mathrm{DR})$ | 10000 |
| Temperature range | -10℃/+40℃ |
| Fraction $p_{\mathrm{LC}}$ | 0,8 |
| Humidity Class | CH |
| Safe overload | 150% of $E_{\mathrm{max}}$ |
| Recommended excitation | 12-24V DC |
| Excitation maximum | 30V DC |
| Transducer material | Stainless steel |
| Atmospheric protection | Hermetically welded |
| Electromagnetic environment class | E2 |
| Number of counts for $E_{\mathrm{max}}$ | ≥Y*5/pLC |
| Software identification | Described in the certificate TC8039 |
Remark:
- The characteristics for $ n_{\mathrm{max}} $ , Y and Z can be reduced separately.
Each load cell produced is provided with an accompanying document with information about its characteristics.
OIML Certificate
OIML Member State The Netherlands
Number R60/2021-A-NL1-26.09 revision 1
Project number 3940755
Page 3 of 3
The above identified Type (represented by the sample(s) identified in the OIML Test Report) have been found to comply with the additional national requirements established by the United States of America (NIST Handbook 44 and NCWM Publication 14), included in the Utilizer Declaration:
R 60 OIML-CS rev.2 Additional requirements from the United States Accuracy class III L;
R 60 OIML-CS rev.2 Additional requirements from the United States Marking requirements
Revision History
| Revision | Date | Change(s) |
| 0 | 2026-06-15 | Initial issue. |
| 1 | 2026-06-29 | Test report NMi-3940755-04 replaced due to correction in it. |