R60/2021-A-NL1-24.20 Rev. 3
OIML Certificate
OIML Member State The Netherlands
Number R60/2021-A-NL1-24.20 revision 3 Project number 3657711 Page 1 of 3
Issuing authority
NMi Certin B.V. Person responsible: M.Ph.D. Schmidt
Applicant and Manufacturer
VPG Precision Transducers India Ltd.
OZ-22 Hi-Tech SEZ
Kancheepuram 602105
Tamil Nadu
India
Identification of the certified type
A single point load cell, with strain gauges. Registered trade name VPG
: VPG force sensors or Tedea- Huntleigh
Type
: vLite 0022, vLite 0031, vLite 0038, vLite 0042
Characteristics
See next page
This OIML Certificate is issued under scheme A.
This Certificate attests the conformity of the above identified Type (represented by the sample(s) identified in the OIML Test Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):
OIML R 60-1:2021 for accuracy class C
This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation above-identified. This Certificate does not bestow any form of legal international approval.
Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.
Issuing Authority
NMi Certin B.V., OIML Issuing Authority NL1 24 March 2025
Certification Board
This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.
OIML Certificate
OIML Member State The Netherlands
Number R60/2021-A-NL1-24.20 revision 3 Project number 3657711 Page 2 of 3
The conformity was established by the results of tests and examinations provided in the associated reports:
No. NMi-3657711-02 R1 dated 20 August 2024 that includes 51 pages;
No. NMi-3657711-03 R1 dated 20 August 2024 that includes 46 pages;
- No. NMi-3657711-04 dated 20 March 2025 that includes 26 pages.
Characteristics of the load cell:
| Characterization of load cell capabilities | Analog-passive load cell | ||
| Maximum capacity ($E_{\text{max}}$) | 15kg up to and including 50kg | 50kg up to and including 100kg | 100kg up to and including 300kg |
| Minimum dead load | 0kg | ||
| Accuracy Class | C | ||
| Rated Output | 2mV/V±0.2mV/V | ||
| Maximum number of load cell intervals(n)^{(1)}$ | 3000 | ||
| Ratio of minimum LC Verification interval(1) $Y=E_{\text{max}}/v_{\text{min}}$ | 14500 | 12000 | 25000 |
| Ratio of minimum dead load output return(1) $Z=E_{\text{max}}/(2*\text{DR})$ | 3000 | 6000 | |
| Input impedance | 410$\Omega$±25$\Omega$ | ||
| Temperature range | -10℃/+40℃ | ||
| Fraction $p_{\text{LC}}$ | 0.7 | ||
| Humidity Class | CH | ||
| Safe overload | 150% of $E_{\text{max}}$ | ||
| Output impedance | 354$\Omega$±5$\Omega$ | ||
| Recommended excitation | 10V AC/DC | ||
| Excitation maximum | 15V AC/DC | ||
| Transducer material | (Coated) Aluminium | ||
| Atmospheric protection | Silicone Rubber | ||
Remarks:
- The characteristics for $ n_{\mathrm{max}} $ , Y and Z can be reduced separately.
Each load cell produced is provided with an accompanying document with information about its characteristics.
OIML Certificate
OIML Member State The Netherlands
Number R60/2021-A-NL1-24.20 revision 3 Project number 3657711 Page 3 of 3
The above identified Type (represented by the sample(s) identified in the OIML Test Report) have been found to comply with the additional national requirements established by the United States of America (NIST Handbook 44 and NCWM Publication 14), included in the Utilizer Declaration:
R 60 OIML-CS rev.2 Additional requirements from the United States Accuracy class III L;
R 60 OIML-CS rev.2 Additional requirements from the United States Marking requirements.
Revision History
This revision replaces the previous versions.
| Revision | Date | Changes |
| 0 | 2024-07-11 | Initial issue. |
| 1 | 2024-08-20 | Corrections and updating Y value. |
| 2 | 2025-03-20 | Addition of 15kg capacity. |
| 3 | 2025-03-24 | Typo correction OIML recommendation number in header |