Skip to content

← R60

R60/2021-A-NL1-24.10 Rev. 0

Applicant
Minebea Intec GmbH
Issuer
NL1
Issuing year
2024
Status
Valid

Source PDF (certificates/R60/2021/r60-2021-nl1-2024-10-rev0.pdf) · extracted by OCR from the digitalized tier of the corpus.

R60/2021-A-NL1-24.10 Rev. 0

OIML Certificate

OIML Member State The Netherlands

Number R60/2021-A-NL1-24.10 revision 0 Project number 3725495 Page 1 of 3

Issuing authority NMi Certin B.V. Person responsible: M.Ph.D. Schmidt

Applicant and Manufacturer

Minebea Intec GmbH Meiendorfer Strasse 205 A D-22145 Hamburg Germany

Identification of the certified type

A compression load cell, with strain gauges, equipped with electronics. Registered trade name : Minebea Intec GmbH

Type

: PR 6226

Characteristics

See next page

This OIML Certificate is issued under scheme A.

This Certificate attests the conformity of the above identified Type (represented by the sample(s) identified in the OIML Test Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

OIML R 60-1:2021 for accuracy class C

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation above-identified. This Certificate does not bestow any form of legal international approval.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Test Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 20 August 2024

Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

OIML Certificate

OIML Member State The Netherlands

Number R60/2021-A-NL1-24.10 revision 0 Project number 3725495 Page 2 of 3

The conformity was established by the results of tests and examinations provided in the associated reports:

Page 2 of 3

No. NMi-3725495-01 dated 20 August 2024 that includes 16 pages;

No. NMi-3725495-02 dated 20 August 2024 that includes 61 pages;

  • No. NMi-3725495-04 dated 20 August 2024 that includes 16 pages.

Characteristics of the load cell:

Characterization of load cell capabilitiesDigital load cell with data processing
Maximum capacity $ E_{\mathrm{max}} $20t up to and including 100t
Minimum dead load0t
Accuracy ClassC
Maximum number of load cell intervals(n)^{(1)}$4000
Ratio of minimum LC Verification interval $ ^{(1)} $ $ Y=E_{\mathrm{max}}/v_{\mathrm{min}} $20000
Ratio of minimum dead load output return $ ^{(1)} $ $ Z=E_{\mathrm{max}}/(2*\mathrm{DR}) $4000
Temperature range-10℃/+40℃
Fraction $ p_{\mathrm{LC}} $0,8
Humidity ClassCH
Safe overload150% of $ E_{\mathrm{max}} $
Recommended excitation24V DC
Excitation maximum28V DC
Transducer materialStainless steel
Atmospheric protectionHermetically welded
Number of counts for $ E_{\mathrm{max}} $≥Y*5/pLC
Software identificationVersion number:01.02.xx^{(2)}$

Remarks:

  1. The characteristics for $ n_{\mathrm{max}} $ , Y and Z can be reduced separately.

  2. xx is a number between 00 and 99 representing updates of the non legally relevant part of the software.

Each load cell produced is provided with an accompanying document with information about its characteristics.

OIML Certificate

OIML Member State The Netherlands

Number R60/2021-A-NL1-24.10 revision 0 Project number 3725495

Page 3 of 3

The above identified Type (represented by the sample(s) identified in the OIML Test Report) have been found to comply with the additional national requirements established by the United States of America (NIST Handbook 44 and NCWM Publication 14), included in the Utilizer Declaration:

  • R 60 OIML-CS rev.2 Additional requirements from the United States Accuracy class III L;

  • R 60 OIML-CS rev.2 Additional requirements from the United States Marking requirements.

Revision History

RevisionDateChange(s)
02024-08-20Initial issue.