R51/2006-A-DK2-24.02
OIML Member State Denmark
OIML Certificate No.
R51/2006-A-DK2-2024.02
OIML CERTIFICATE ISSUED UNDER SCHEME A
| OIML Issuing Authority | |
| Name: FORCE Certification A/S Address: Park Allé 345, 2605 Brøndby, Denmark Person responsible: Per Rafn Crety | |
| Applicant | |
| Name: Shanghai Teraoka Electronic Co., Ltd. Address: No. 6058 of Nanting Road, Tinglin town, Jinshan District Shanghai CHINA | |
| Manufacturer: Shanghai Teraoka Electronic Co., Ltd. | |
| Identification of the certified type (the detailed characteristics will be defined in the additional pages) XCK-30/XCK-60 | |
| Designation of the module (if applicable) Automatic Catchweighing instrument | |
| This OIML Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML type evaluation report) with the requirements of the following Recommendation of the International Organization of Legal Metrology(OIML): OIML R 51-1, Edition(year):2006 For accuracy class(if applicable):XIII or Y(a) |
| Revision No. | Date | Description of the modification |
|---|---|---|
| Initial version | 09 August 2024 | - |
Descriptive annex
Characteristics
Accuracy class: X(III), Y(a)
Weighing range: Single-interval
Weighing mode: Dynamic
Maximum capacity (Max): 30 and 60 kg
Minimum capacity (Min): $ \geq50\times e $
Verification scale interval (e=): $ \geq10\mathrm{g} $
Number of verification scale intervals (n): $ \leq 3 0 0 0 $
Belt speed: 30 m/min
Warm-up time: None
Temperature range: -10 $ ^{\circ} \mathrm{C} $ to $ +40 , \mathrm{^\circ C} $
Supply voltage: 230 VAC, 50/60 Hz
Electromagnetic class: E2
Humidity: Non-condensing
Software
Identification of the software version is performed during power-up.
The approved software version is v4.xx, where xx can be 00-99 and represents minor non-legal changes.
Devices
Initial zero setting device ( $ \leq20% $ of Max)
Automatic zero setting device ( $ \leq4% $ of Max)-dynamic mode only
Semiautomatic zero setting device ( $ \leq4% $ of Max)-static mode only
Zero tracking device ( $ \leq4% $ of Max)-static mode only
No motion detection and indication
Detection of significant fault
Interfaces
Ethernet for communication to peripherals.
USB (for setup of the instrument only)
The interface does not have to be secured.