R51/2006-A-DK2-23.03

OIML Certificate No.

R51/2006-A-DK2-2023.03

OIML CERTIFICATE ISSUED UNDER SCHEME A

OIML Issuing Authority
Name: FORCE Certification A/S
Address: Park Allé 345, 2605 Brøndby, Denmark
Person responsible: Per Rafn Crety
Applicant
Name: Marel Iceland ehf.
Address: Austurhraun 9210 Gardabear Iceland
Manufacturer: Marel Iceland ehf.
Identification of the certified type (the detailed characteristics will be defined in the additional pages)
M2400-IWU-xLC
Designation of the module (if applicable)
Automatic catchweighing weighing instrument
This OIML Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML type evaluation report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):
Jens Hovgård Jensen
2023-11-16
OIML R 51-1, Edition (year): 2006
For accuracy class (if applicable): Y(a)
OIML Certificate No.R51/2006-A-DK2-2023.03
This OIML Certificate relates only to metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML Recommendation identified aboveThis OIML Certificate does not bestow any form of legal international approval.
The conformity was established by the results of tests and examinations provided in the associated OIML reports:Type examination report:No.121-35487.10,dated01August2023,that includes66 pagesType evaluation report:No.121-35487.90.20,dated09October2023,that includes19 pages
The technical documentation relating to the identified type is contained in documentation file:121-35487OIML Certificate History
Revision No.DateDescription of the modification
Initial version16November2023-
Identification, signature and stampThe OIML Issuing AuthorityFORCE CertificationA/SDate:16November2023Jens Hovgård JensenCertification Manager
Important note: Apart from the mention of the Certificate's reference number and the name of theOIML Member State in which the Certificate is issued,partial quotation of theCertificate and of the associated OIML type evaluation report(s) is not permitted,although either may be reproduced in full.

Descriptive annex

Description

The automatic catch-weighing instrument is a belt weighing scale designated M2400-IWU-xLC where x represents the number of load cells in the instrument and is intended for dynamic weighing. It consists of an electronic weighing indicator M2400 and a load receptor with 1 or 2 load cells (often designated Inline Weigher) or 4 load cells (often designated MaxiweigherII). The instrument can operate as a price labeller.

Technical Data

  • Accuracy class: Y(a)

  • Weighing range: Single-interval

  • Weighing mode: Dynamic

  • Maximum capacity (Max): 10 kg to 100 kg

  • Minimum capacity (Min): $ \geq2 0 \times\mathrm{e} $

  • Verification scale interval (e=): $ \geq10\mathrm{g} $

  • Number of verification scale intervals (n): $ \leq 3 0 0 0 $

  • Maximum preset tare: 100 % of Max

  • Belt speed: $ \leq 0. 8 \mathrm{~ m / s} $

  • Warm-up time:

  • Maximum time between automatic zero-setting: 28 minutes

  • Temperature range: -10 $ ^{\circ} \mathrm{C} $ to $ +40 , \mathrm{^\circ C} $

  • Supply voltage: 230 VAC, 50/60 Hz

  • Electromagnetic class: E2

  • Humidity: Non-condensing

cation

Technical data, indicator:

  • Apportionment factor: 0.25 $ \mu\mathrm{V} $

  • Exitation voltage: ±3VDC bipolar (6V effective)

  • Maximum input impedance: 87 ohm

  • Minimum input impedance: 1100 ohm

Software

The firmware of M2400 is displayed during start up and can also be found in the Main menu/Identity.

The software version is given on the form a.bb-cc where a reflects major changes including the legal software, bb reflects major bug fixes not affecting the legal software and cc reflects minor bug fixes.

Approved firmware versions are 1.01-00 or higher.

The firmware also lists the version of the weighing module. Approved version is 100.

Devices

  • Power up test

  • Initial zero setting device (max. 20 % of Maximum capacity)

  • Semi-automatic zero setting device (max. 4 % of Maximum capacity)

  • Automatic zero setting device (max. 4 % of Maximum capacity)

  • Zero tracking device (max. 4 % of Maximum capacity)

  • Zero annunciator

  • Preset tare device

  • Storing preset tare

  • Preset Tare indication

  • Tare (NET) annunciator

  • Stable annunciator for empty belt

  • Extended indicating device (service mode only)

  • Gravity compensation

  • Detection of significant fault

  • DSD built in data storage device (Alibi)

  • Transmission of measurement data to an external data storage device (Alibi), Marel M12-DSD or another certified external alibi storage device

  • Transmission of measurement data to an external "Marel Secure Print" application software (Evaluation Certificate No. DK0199-16.07)

  • Price and Weight labelling when used with Marel Secure Print

Interfaces

  • RS232 (2x)

  • CAN

  • USB

  • Ethernet

Load cells

Inline Weigher models with one load cell (1LC model) or two load cells (2LC model): Flintec PCB C3

MaxiweigherII with 4 (4LC model) load cells:

Flintec SB6 C3 or other certified load cells with same or better specifications

Sealing Measures

There are two alternative methods for sealing the load cells to the indicator.

  • Either:

The load cell serial numbers are stored in a CON protected memory in M2400, the serial numbers can be viewed under Menu / Audit and should be identical to those marked on the load cells,

Or:

access to the load cell connection is physically prevented by sealing the M2400 indicator enclosure with a tamperproof sticker or a wire and plumb seal.

On load receptors using four load cells, the junction box, TB-4, is always sealed by using a tamperproof sticker placed on one of its sides.

There are two alternative methods to protecting the legally relevant parameters.

  • Either:

The instrument is secured by two event counters, CAL and CON, which are incremented each time the calibration or sealed configuration parameters are changed. At verification the value of the two event counters is written on a brittle plastic sticker - sealed with a verification mark - next to it. If the value of the CAL or CON differs from the one written at verification time, the seal is broken.

Or:

the access control is set to "Locked" either by switch the DIP switch S2-SEAL to the "ON" position, or by setting the Access control in menu Login to "Locked". When Access control is set to "Locked" no changes can be done to CON or CAL parameters. The Access level "Locked" can only be changed by accessing the S1 push bottom on the PCB. By sealing the M2400 indicator enclosure with a tamperproof sticker or a wire and plumb seal the access level cannot be adjusted without breaking a seal.

Item in figure 17DescriptionFunction
1M2400Enclosure
2Display areaMain display - user interface
3Main screenMain screen
4Programmable function keyFour programmable function keys for operating the indicator key
5Manufacturing logoMarel logo, Indicator type and place of manufacturing
6Place for markers and stickers
7KeypadsNumerical 0 to 9 and the decimal point. Used for entering numerical data.
8Page keyTo cycle between pages and as a page back key.
9Up and down keyUsed for navigating menu items and marking selections.
10Check keyThe check key is used to enter menu and confirming data selection.
11Rating plateIndicator rating plate
12Cable glands

Figure 1: Indicator front view

Item in figure 18DescriptionFunction
1Status barUsed for running indication, service level when logged in and error symbols
2Product listList of last seven products weighed. Indicates weight, quality, length, result (whether weighing has been successful) and alibi (unique ID for each product)
3Program nameName of any user grading program if active
4Pre-set tareShows current pre-set tare.
5CapacityCurrent set capacity: Max/Min/e=d and verification scale intervals. Can never be larger than marked on load receptor.
6Start/StopIn picture - Stop symbol is shown (if stopped start)
7StatusREADY, if running and ready to weigh STOPPED if belt not running.
8Piece rateLeft most, calculated current piece rate (units/min) Right most, recommended piece rate (units/min) for current settings
9Max Item lengthRecommended maximum length of product for current settings.
10Measured speedMeasured current belt speed (m/s)
11Nominal speedUser setpoint for belt speed (m/s)
12Primary weight displayPrimary weight display.
13Service levelFunction key, when pressed will take user to service level login page. Password: 62735.

Figure 2: Main screen of user interface

Figure 3: Layout and dimensions of Inline Weigher, 1 or 2 load cells

Belt HeightRange of Adjustable legs
800+/-100
850+/-100
900+/-100
1000+/-100
1100+/-100
1200+/-100

NOTE: all dimensions are in millimetres (mm)

Figure 4: Layout and dimensions of MaxiweigherII, 4 load cells

Figure 6: Sealed TB-4 box