R51/2006-A-DK2-22.01
| OIML Certification System | FORCE Certification |
| OIML Member State Denmark | OIML Certificate No. R51/2006-A-DK2-2022.01 |
| OIML CERTIFICATE ISSUED UNDER SCHEME A | |
| OIML Issuing Authority Name: FORCE Certification A/S Address: Park Allé 345, 2605 Brøndby, Denmark Person responsible: Per Rafn Crety | |
| Applicant Name: Flintec UK Ltd. Address: Caxton House, Caxton Place, Pentwyn Cardiff CF23 8HG, United Kingdom | |
| Manufacturer Flintec UK Ltd. | |
| Identification of the certified type (the detailed characteristics will be defined in the additional pages) ER500-C | |
| Designation of the module (if applicable) Weighing transmitter for automatic catchweigher | |
| This OIML Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML type evaluation report) with the requirements of the following Recommendation of the International Organization of Legal Metrology(OIML): OIML R 51-1, Edition(year):2006 For accuracy class(if applicable):Y(a),XIII,Y(b)andXIIII | |
| Revision No. | Date | Description of the modification |
|---|---|---|
| Initial version | 16 June 2022 | - |
Descriptive annex
Characteristics
Type: ER500-C
Accuracy class: Y(a), XIII, Y(b) and IIII
Weighing range: Single-interval, multi-range or multi-interval
Maximum number of Verification Scale Intervals: 10000 (class Y(a) and XIII), 1000 (class Y(b) and XIIII) or 3×10000 (class Y(a) and XIII), 3×1000 (class Y(b) and XIIII)
Maximum tare effect: -Max for single-interval and multi-range
- $ \mathrm{M a x}_{1} $ for multi-interval
Fractional factor: $ p^{\prime} i=0.5 $
Minimum input-voltage per VSI: 0.4 $ \mu\mathrm{V} $
Extra warm-up time for $ 0.4 \mu\mathrm{V} \leq e < 0.8 \mu\mathrm{V} $: 18 minutes
Extra warm-up time for $ 0.8 \mu\mathrm{V} \leq e $: 3 minutes
Maximum time between automatic zero-setting: 150 minutes
Excitation voltage: 5 VDC
Load cell interface: 4-wire or 6-wire
Minimum input-impedance: 43 ohm
Maximum input-impedance: 1200 ohm
Mains power supply: 9-32 VDC - not to be supplied from DC mains.
Operational temperature: -15 $ ^{\circ} \mathrm{C} $ to +55 $ ^{\circ} \mathrm{C} $
Maximum cable length between ER500-C and junction box 1533 m/mm $ ^{2} $
Software
The software version is displayed during the start-up of the indicator. (Alternating with the TAC number). The version format is xx.yy.zz, where x is the basic software family, while yy is version numbers for minor legally relevant changes and zz is changes and corrections not influencing the legal function of the software.
The approved software version is 01.01.zz.
TAC number
The non-resettable Traceable Access Code is displayed during the start-up of the indicator in the format: xxxxx. (Alternating with the software version).
Devices
Initial zero setting device ( $ \leq 20% $ of Max)
Semi-automatic zero setting device ( $ \leq 4 % $ of Max)
Zero tracking device ( $ \leq 4 % $ of Max)
Semi-automatic subtractive tare balancing device
Units (Allowed units are g, kg and t.)
Stable equilibrium, Zero and Net indicators.
Display and buttoms
If the ER500-C is equipped with a display and buttoms these are are for set-up and service purpose only.
Interfaces
- RS232
RS-485/RS-422
USB
Ethernet
3 logical inputs
3 logical outputs
Analogue Output
The interfaces do not have to be secured.
Sealing
Access to the set-up and calibration facility requires that a calibration jumper is removed from the main board. The jumper can be accessed from the outside, top part of the housing.
The indicator also has a Traceable Access Counter, which increment each time the calibration or legal part of the set-up has been changed.
The sealing of the calibration jumper, which also prevents the housing from being dismantled - is accomplished with a brittle plastic sticker. The sticker is placed across the opening designated 30 behind which the calibration jumper is located.
Figure 2 Front panel display with LED indicators and buttoms.
Figure 3 Top side label of ER500-C.