Skip to content

← R139

R139/2018-A-NL1-24.03 Rev. 0

Applicant
Emerson Process Management Flow B.V.
Issuer
NL1
Issuing year
2024
Status
Valid

Source PDF (certificates/R139/2018/r139-2018-nl1-2024-03-rev0.pdf) · extracted by OCR from the digitalized tier of the corpus.

R139/2018-A-NL1-24.03 Rev. 0

OIML Certificate

OIML Member State The Netherlands

Number R 139/2018-A-NL1-24.03 revision 0 Project number 3800929 Page 1 of 4

Issuing authority NMi Certin B.V. Person responsible: M.Ph.D. Schmidt

Applicant Emerson Process Management Flow B.V. Neonstraat 1 6718 WX Ede The Netherlands

Manufacturer Micro Motion, Inc. 7070 Winchester Circle Boulder, Colorado 80301 United States of America

Identification of the certified type A measuring device (Coriolis), for the measurement of Compressed Natural Gas (CNG). Manufacturers mark: Micro Motion, Inc. Type: CNG 050

Characteristics See following page(s)

This OIML Certificate is issued under scheme A.

This Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML Type Evaluation Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

R 139:2018 "Compressed gaseous fuel measuring systems for vehicles"

Accuracy class 1,5

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation identified above. This Certificate does not bestow any form of legal international approval.

This certificate and supporting reports comply with the requirements of OIML-CS-PD-07 clause 6.2.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Type Evaluation Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 22 July 2024

Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

This document is digitally signed and sealed. The digital signature can be verified in the blue ribbon at the top of the electronic version of this certificate.

OIML Certificate

Number R 139/2018-A-NL1-24.03 revision 0

Project number 3800929

Page 2 of 4

The conformity was established by the results of tests and examinations provided in the associated reports:

For the measuring device and flow transmitter, brand Micro Motion Inc., type CNG 050 and type MVD 700 and MVD 800:

  • No. NMi-3800929-01 dated 22 July 2024 that includes 28 pages.

Characteristics of the measuring device

In Table 1, the general characteristics of the measuring device are presented.

The construction of the measuring device is recorded in the Documentation folder no. TC11012-1, and TC7057-17 for the MVD core processors.

Measuring device "CNG 050"

Table 1 General characteristics

Minimum-maximum flow rate1,3-77kg/min;
Minimum measured quantity1kg
Maximum pressure317 or 345 bar(g)(depending on flange type)
Environmental classesM2/E2
Ambient temperature range-40-+55℃;condensing humidity
Product temperature range-25-+55℃
Intended for the measurement ofCompressed Natural Gas
Power supply voltage24VDC±20%(with MVD Direct Connect I.S)15-26VDC

Flow transmitter "MVD 700 Core Processor or MVD 800 Enhanced Core Processor" Table 2 General characteristics

Environmental classesM2/E2
Ambient temperature range-40-+55℃; condensing humidity
Power supply voltage24VDC

OIML Certificate

OIML Member State The Netherlands

Number R 139/2018-A-NL1-24.03 revision 0

Project number 3800929

Page 3 of 4

Table 3 Software versions of the MVD series electronics

VersionChecksumVersionChecksumVersionChecksum
700 Core Processor
2.051FF2.7F666*)3.218D0
2.12B3F2.81DEA*)3.3B0D1
2.290053.0D00D*)3.4073A9
2.3D75B3.0-ETO1715397D6*)3.42F00C
2.4474F3.11-ETO1941314AD3.5011AA
2.514AD3.121F1B*)3.523C4A
2.6D732*)3.13-ETO189518BF8*)
800 Enhanced Core Processor
3.11891378AB3.9447EB3E104.80F1583A44
3.219893B9993.96756C1BFD4.96083BF9B
3.30A73D25DA4.00C582F8435.084D368E71
3.427FA82CE84.028D61C3685.1082C541D9
3.50D9343F054.1440860C635.20BD69FDD6
3.52132CCB634.202983A9BE5.22F4A8D922
3.6A9CA4E814.21-ETO21931D63492595.23-ETO45214B1D70450
3.61-ETO171709AA358FF4.40B280233F5.3065828884
3.7BE73CD624.42D7BA08415.33BF3164F6
3.71-ETO18982580D32B64.506B48C6245.400218C30B
3.88CA8E7D14.51-ETO32353BC1660E85.50761BF30F
3.81-ETO207757931CE3D4.51-ETO33244D7B811355.605BE64A27
3.958CB3E0C4.60DDB76E3C
3.91-ETO2115665F98DD74.70AEB92E3F

Notes:

*) Indicated as 14AD, corrected from version 3.2

OIML Certificate

OIML Member State The Netherlands

Number R 139/2018-A-NL1-24.03 revision 0 Project number 3800929

Page 4 of 4

Installation conditions:

Installation of the measuring device with the MVD Direct Connect barrier is preferred to protect against EMI.

Certificate history:

This revision replaces the previous version.

RevisionDateDescription of the modification
Initial22 July 2024First issue.