R139/2018-A-NL1-22.07 Rev. 1

Applicant
Cetil Dispensing Technology, S.L.
Issuer
NL1
Issuing year
2026
Status
Valid

Source PDF (certificates/R139/2018/r139-2018-nl1-2022-07-rev1.pdf) · extracted by OCR from the digitalized tier of the corpus.

R139/2018-A-NL1-22.07 Rev. 1

OIML Certificate

OIML Member State The Netherlands

Number R139/2018-A-NL1-22.07 revision 1

Project number 3961338

Page 1 of 4

Issuing authority NMi Certin B.V. Person responsible: M.Ph.D. Schmidt

Applicant and Manufacturer Cetil Dispensing Technology, S.L. C/Pelaya 37 Polígono Industrial Río de Janeiro 28110 Algete (Madrid) Spain

Identification of the compressed gas (CG) dispenser for hydrogen certified type Manufacturers mark: Cetil Type: E30 H2

Characteristics See following page(s)

This OIML Certificate is issued under scheme A.

This Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML Type Evaluation Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

R 139:2018 "Compressed gaseous fuel measuring systems for vehicles"

Accuracy class

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation identified above. This Certificate does not bestow any form of legal international approval.

This certificate and supporting reports comply with the requirements of OIML-CS-PD-07 clause 6.2.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Type Evaluation Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 6 February 2026

Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

This document is digitally signed and sealed. The digital signature can be verified in the blue ribbon at the top of the electronic version of this certificate.

OIML Certificate

Number R139/2018-A-NL1-22.07 revision 1 Project number 3961338 Page 2 of 4

The conformity was established by the results of tests and examinations provided in the associated report(s):

  • No. NMi-2546086-01 dated 5 August 2022 that includes 28 pages.

  • No. NMi-3961338-01 dated 6 February 2026 that includes 15 pages.

Characteristics of the measuring instrument

In Table 1 the general characteristics of the measuring instrument are presented.

The construction of the measuring instrument is recorded in the Documentation folder no. T9010-1.

Table 1 General characteristics

Accuracy class2
Minimum-maximum flow rateDepending on the flow sensor used, see the relevant OIML certificate
Ratio $Q_{\mathrm{max}}/Q_{\mathrm{min}}$≥10
Minimum measured quantity1kg
Maximum pressureDepending on the flow sensor used, see the relevant OIML certificate.
Environmental classesDepending on the parts used, see the relevant OIML certificates.
Ambient temperature rangeDepending on the parts used, see the relevant OIML certificates.
Product temperature rangeDepending on the flow sensor used, see the relevant OIML certificate.
Intended for the measurement ofHydrogen($\mathrm{H}_{2}$)

Each CG dispenser consists at least of one measuring system made up of the following essential parts: A flow meter (measurement sensor and transducer); and

  • A calculating/indicating device.

The same housing of the dispenser can comprise of one or more measuring systems. When more than one measuring systems are in one housing, one calculating/indicating device may be a common part of the measuring systems.

To correct the amount of hydrogen delivered, the electronic calculating and indicating device takes into consideration that the dispenser is an empty hose measuring system and also the pressure, temperature and volume of relevant sections of the dispenser. Pressure and temperature of relevant sections are measured via transmitters, and the volume is configured in the calculating/indicating device. The monitorization of the pressure and temperature allows stopping or even not starting a delivery if out of range values according to the fuelling protocol are measured and this also guarantees the integrity of the measurement.

In Table 2 the overview of the essential parts of the measuring instrument are presented.

OIML Certificate

OIML Member State The Netherlands

Number R139/2018-A-NL1-22.07 revision 1

Project number 3961338

Page 3 of 4

Table 2 Overview parts of the measuring instrument

PartProducerTypeOIML certificateOIML ReportsRemarks
Measurement sensorEmerson Process Management Flow B.V.HPC0xxR139/2018-A-NL1-21.03-With MVD electronics as mentioned in the OIML certificate.
Measurement sensorRheonikRHM04 or RHM10R139-2018-A-NL1-22-01-To be used with Rheonik transducer.
Measurement transducerRheonikRHE42 or RHE45R139-2018-A-NL1-22.04-To be used with Rheonik sensor.
Calculating / indicating deviceCetilEAS2-See Table 4-

Table 3 General characteristics of the calculating/indicating device type EAS2

ProducerCetil
TypeEAS2
Documentation folderTC8491-3
Reports-No.NMi-15200617-01 dated 14 April 2016 that includes 85 pages.-No.NMi-15200617-02 dated 14 April 2016 that includes 20 pages.-No.NMi-15200617-04 dated 14 April 2016 that includes 34 pages.
Maximum volume indication8 positions 99999999
Maximum unit price8 positions 99999999
Maximum price to pay8 positions 99999999
Environmental classesM3/E2/H3
Ambient temperature range-40℃/+55℃
Software identificationSee below table
Power supply voltage230 AC;50Hz

The validated software versions and checksums are:

Software versionChecksum for module
MCONMVISMEESM420
02-0064465ED19119A678458C81157305BF75
02-05E975D8D17ABE6DD0B0D247411D19E9B9

OIML Certificate

OIML Member State The Netherlands

Number R139/2018-A-NL1-22.07 revision 1 Project number 3961338

Page 4 of 4

Certificate history:

This revision replaces the previous version.

RevisionDateDescription of the modification
05 October 2022Initial revision
16 February 2026Addition of test data and update to new template