R139/2018-B-DK2-20.02
| OIML Certification System | FORCE Certification |
| OIML Member State Denmark | OIML Certificate No. R139/2018-B-DK2-2020.02 |
| OIML CERTIFICATE ISSUED UNDER SCHEME B | |
| OIML Issuing Authority Name: FORCE Certification A/S Address: Park Allé 345, 2605 Brøndby Denmark Person responsible: Kurt Rasmussen | |
| Applicant Name: Heinrichs Messtechnik GmbH Address: Robert-Perthel-Strasse 9, D 50739 Köln | |
| Manufacturer Name: Heinrichs Messtechnik GmbH Address: Robert-Perthel-Strasse 9, D 50739 Köln | |
| Identification of the certified type (the detailed characteristics will be defined in the additional pages) A Coriolis meter Type TMU-W004 with a transmitter type UMC4 or UMC4-RM | |
| Designation of the module (if applicable) Coriolis measuring device for measurement of Hydrogen to be used in a hydrogen refuelling dispenser. | |
| This OIML Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML type evaluation report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML): OIML R 139-1, Edition (year): 2018 For accuracy class (if applicable): 2 | |
| OIML Certificate No.R139/2018-B-DK2-2020.02 | ||
| This OIML Certificate relates only to metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML Recommendation identified aboveThis OIML Certificate does not bestow any form of legal international approval. | ||
| The conformity was established by the results of tests and examinations provided in the associated OIML type evaluation reportNo.120-25992.01.02_1 that includes 59 pages | ||
| The technical documentation relating to the identified type is contained in documentation fileTask no.120-25992OIML Certificate History | ||
| Revision No. | Date | Description of the modification |
| Revision 0 | 17-12-2020 | Original certificate |
| Identification, signature and stampThe OIML Issuing AuthorityDate:17-12-2020Michael Møller NielsenCertification manager | ||
| Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate is issued, partial quotation of the Certificate and of the associated OIML type evaluation report(s) is not permitted, although either may be reproduced in full. | ||
Descriptive annex
Flow tests is done according to OIML R139:2018-2 table 9 for meters which do not include determination of the actual flowrate range but the dispensed mass. Therefore, the actual flowrate is not determined and verified.
All flow tests are done with the meter installed in an HRS at a stable temperature of $ - 4 0^{\circ} \mathrm{C}. $
Essential parts:
Flow sensor:
Flow transmitter:
Manufacture: Heinrichs messtechnik GmbH
Ambient temperature: -40...+55 $ ^{\circ} \mathrm{C} $
Mains DC voltage: 19-36 VDC
Software version: 5.241
Output: Pulses
Or:
Flow transmitter:
Environmental class: M2
Ambient temperature: -40...+55 $ ^{\circ} \mathrm{C} $
Mains DC voltage: 19-36 VDC
Software version: 5.241
CRC Checksum: D2D5
Output: Pulses
Sealing:
The flow sensor and the transmitter are sealed by a physical seal according to manufacture sealing plan.
Sealing of the transmitter UMC4:
Sealing after commissioning:
Sealing with sticker
Sealing of the transmitter UMC4-RM:
Sealing with wire and lead
Rack mounted enclosure
Sealing of the rack mounted transmitter
Sealing of the "custody stamp" Switch:
Page 4 of 6 pages
Identification plate:
The sensor and transmitter, shall bear a permanent, non-transferable, and easily readable identification plate or label giving the following information:
manufacturer's trade mark/corporate name;
year of manufacture;
type designation / model number;
- type approval number and (area allowed for) verification marks, according to national legislation;
The following metrological and technical characteristics, where applicable, shall be provided either on the identification plate, or may be visible either permanently, or on demand on the indicating device, as appropriate:
measuring range (minimum flow rate, $ Q_{\mathrm{min}} $ , and maximum flow rate, $ Q_{\mathrm{max}} $ );
maximum pressure of the gas, $ P_{\mathrm{max}} $;
type of the gas be measured (hydrogen);
maximum temperature of the gas, $ T_{\mathrm{max}} $;
minimum temperature of the gas, $ T_{\mathrm{min}}. $
ambient temperature range;
the applicable environmental class M2.
Example of an identification plate(s).
Sensor TMU-W004
KOBOLD Group
D-50739 Köln
Robert-Perthel-Str. 9
0158
DMT 01 ATEX E 149 X
II 1/2 G Ex ia IIC T6-T2 Ga/Gb
IECEx BVS 11.0084 X
IEx ia IIC T6-T2 Ga/Gb
EXCITER CIRCUIT TYP EC1
Type: TMU-W004-4500-A00-JL-10-U-H
| Ser. No.: | 350090 |
| TAG No.: | OIML R139-1 |
| MF-Date: | 2020/04 |
| Wetted Parts: | 1.4571 |
| Medium Temp: | -40°C to 55°C |
| Tamb: | -40°C to +55°C |
| Connection: | 9/16" - 18 UNF |
| PS: 1050 bar | PT: 1600 bar |
| PN Body: | 0,4 bar |
| Qmin = | 8 kg/h |
| Qmax = | 240 kg/h |
| Sensor Constant C: | 757,684 |
| PED: Art.4,Cat.3 | Protection:IP65 |
| Connection: | HAN R23 Conn. |
Custody transfer parts certificate number. XXXXXX
Transmitter UMC4-RM
Type:UMC4-D12A31H
MF-Date: 2020/04
Ser. No.: 351462
KOBOLD Group
TAG No.: OIML R139-2
Supply: 19-36 VDC/4,5 W
D-50739 Köln
Tamb: -20°C to 55°C
Robert-Perthel-Str. 9
Signal-outputs passive Ex [ia Ga] I-OUT 4-20mA/HART, 2x IMP-OUT
Sensor circuits Ex ia Ga
0158
Custody transfer parts certificate number: XXXXXX
See Manual
BVS 20 ATEX E xxx X
! → i
IECEx BVS 10.xxxxxX
II 3G Ex ec [ia Ga] II C T6-T4 Gc