R137/2012-A-NL1-21.05
OIML Certificate
OIML Member State The Netherlands
Number R137/2012-A-NL1-21.05 Project number 2518128 Page 1 of 6
Issuing authority
NMi Certin B.V.
Person responsible: M.Ph.D. Schmidt
Applicant and Manufacturer
SICK Engineering GmbH Bergener Ring 27 D-01458 Ottendorf-Okrilla Germany
Identification of the certified type
An ultrasonic gas meter Type: FLOWSIC600-XT
Characteristics
See page 2 and further
This OIML Certificate is issued under scheme A
This Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML Type Evaluation Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):
R 137-1 (2012) "Gas meters"
Accuracy class
0,5
This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation identified above. This Certificate does not bestow any form of legal international approval.
Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Type Evaluation Report(s) is not permitted, although either may be reproduced in full.
Issuing Authority
NMi Certin B.V., OIML Issuing Authority NL1 4 August 2021
Certification Board
This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.
Reproduction of the complete document only is permitted.
The notification of NMi Certin B.V. as Issuing Authority can be verified at www.oiml.org
OIML Certificate
OIML Member State The Netherlands
Number R137/2012-A-NL1-21.05 Project number 2518128 Page 2 of 6
The conformity was established by the results of tests and examinations provided in the associated report:
- No. NMi-2518128-01 dated 4 August 2021 that includes 64 pages.
Characteristics of the measuring instrument
In Table 1 the general characteristics of the measuring instrument are presented. Table 2 gives an overview of the general characteristics of the family of instruments.
Table 1 General characteristics
| Destined for the measurement of | Gas volume | |||
| Environmental classes | M2/E2 | |||
| Accuracy class | 0,5 | |||
| Minimum operating pressure | atmospheric | |||
| Maximum operating pressure | depending on transducer type, see below | |||
| Ambient temperature range | -40...+70℃ | |||
| Process gas temperature range | -46...+110℃ | |||
| Designed for | Non-condensing humidity&condensing humidity | |||
| Orientation | Horizontal, vertical up and vertical down(all orientations) | |||
| Power supply voltage | 6-24V DC | |||
| Software identification | Version number:1.05.00Checksum:0x78E1 | |||
| Transducers | Type | Frequency[kHz] | pmax[bar] | Beam angle[°] |
| 46 | 205 | 180 | 15,9 | |
| 22 | 205 | 180 | 12,2 | |
| S6 | 205 | 103 | 11,9 | |
| 26 | 205 | 180 | 11,7 | |
| S5 | 350 | 105 | 9,5 | |
| T210 | 205 | 105 | 13,7 | |
| T220 | 205 | 180 | 12,2 | |
| H210 | 205 | 105 | 6,8 | |
| T218 | 205 | 105 | 13,7 | |
| H218 | 205 | 105 | 6,8 | |
| T228 | 205 | 180 | 12,2 | |
OIML Certificate
OIML Member State The Netherlands
Number R137/2012-A-NL1-21.05
Project number 2518128
Page 3 of 6
Table 2 Meter types, path configuration & installation conditions
The meter consists of a cylindrical spool piece with 4 or 8 horizontal direct paths. Multiple configurations can be combined in a single housing according to the table below.
When the meter is equipped with a double meter configuration, the pay and check meter shall have a different indicator. A clear distinction between pay and check indicator shall be present, both meters must have their own nameplate with unique serial number.
| FLOWSIC600-XT... | Metrological relevant transducer paths | Check paths(non-metrological relevant) | Flow disturbance | Minimum installation conditionsup- and downstream[1] |
| Standard | 4 | - | Mild | 2D-FC-8D-USM-3D |
| 2plex | 4 | 1 | Mild | 2D-FC-8D-USM-3D |
| Quatro | 4 | 4 | Mild | 2D-FC-8D-USM-3D |
| Forte | 8 | - | Mild&Severe | 5D-USM-3D |
[1] xD = Minimum straight inlet piping length
FC = Flow conditioner: PTB design, FL600, drawing no. 9211779, d.d. 2014-05-19
USM = Ultrasonic gas meter
OIML Certificate
Number R137/2012-A-NL1-21.05
Project number 2518128
Page 4 of 6
Table 3 General characteristics of the family of instruments
| Standard(4Path),2plex(4Path),Quatro(4Path)和Forte(8Path) | |||||
| Diameter Nominal | Internal bore | Maximum $V_{\mathrm{max}}$ [m/s] | Minimum | ||
| minimum [mm] | maximum [mm] | $V_{\mathrm{t}}$ [m/s] | $V_{\mathrm{min}}$ [m/s] | ||
| 4"(100mm) | 9496 | 110 | 3922 | 254 | 0.31 |
| 6"(150mm) | 142 | 160 | 3922 | 175 | |
| 8"(200mm) | 190 | 210 | 3922 | 157 | |
| 10"(250mm) | 235 | 270 | 3922 | 154 | |
| 12"(300mm) | 270 | 315 | 3784 | 1.5 | |
| 14"(350mm) | 315 | 340 | 3564 | ||
| 16"(400mm) | 360 | 400 | 3548 | ||
| 18"(450mm) | 405 | 450 | 3464 | ||
| 20"(500mm) | 450 | 500 | 3464 | ||
| 24"(600mm) | 495 | 550 | 3464 | ||
| 26"(650mm) | 540 | 600 | 3396 | ||
| 28"(700mm) | 585 | 650 | 3307 | ||
| 30"(750mm) | 630 | 700 | 3208 | ||
| 32"(800mm) | 675 | 750 | 3105 | ||
| 34"(850mm) | 720 | 800 | 2934 | ||
| 36"(900mm) | 765 | 850 | 2840 | ||
| 38"(950mm) | 810 | 900 | 2749 | ||
| 40"(1000mm) | 855 | 950 | 2709 | ||
Remarks regarding table 3
[1] If higher values are chosen for $ Q_{\mathrm{min}} $ or $ Q_{t} $ and/or lower values for $ Q_{\mathrm{max}} $ it has to be taken into account that:
$$ Q _ {\max } / Q _ {\min } \geq 5 0 $$
$$ Q _ {\max } / Q _ {t} \geq 1 0 $$
- If ratio $ Q_{\mathrm{max}} / Q_{\mathrm{min}}\geq 5 $ and < 50 then: $ Q_{\mathrm{max}} / Q_{t}\geq 5 $
[2] Depending on the internal bore, the flow rate in $ m^{3}/h $ can be calculated.
The corresponding flow rates can be calculated as follows:
$$ Q = v \cdot \frac {1}{4} \cdot \pi \cdot D ^ {2} \cdot 3 6 0 0 $$
Where:
Q = flow rate $ [\mathrm{m}^{3} / \mathrm{h}] $
v = velocity [m/s]
D = internal diameter [m]
[3] Example calculation for a 4" ultrasonic gas meter with an internal bore of 94,96 mm: $ Q_{\mathrm{m a x}}=3 9,2 2\cdot\frac{1}{4}\cdot3,1 4 1 5 9\cdot9 4,9 6^{2}\cdot3 6 0 0=1 0 0 0 \mathrm{m}^{3} / \mathrm{h} $
OIML Certificate
OIML Member State The Netherlands
Number R137/2012-A-NL1-21.05 Project number 2518128 Page 5 of 6
Installation conditions:
Installation requirements
Any components which could affect the gas flow must be avoided within the prescribed inlet pipe length. The inlet pipe must be designed as a straight pipe section of the same nominal diameter as the gas meter. The necessary straight pipe length is stated on the name plate of the meter.
The installation requirements, based on accuracy class and mild and/or severe flow disturbance are specified in table 2. The used flow conditioner (if applicable) is of type: "PTB design flow conditioner", type FL600, as documented in drawing no. 9211779, dated 2014-05-19.
Working pressure
The spool piece and the transducers can be used up to the specified pressure as given in table 1.
The two pressures, at which the meter shall be calibrated, without intermediate adjustment over its full flow rate range during initial verification ( $ p_{\mathrm{min}} $ calibration and $ p_{\mathrm{max}} $ calibration),shall be equal to the operating pressure range ( $ p_{\mathrm{min}} $ and $ p_{\mathrm{max}} $ )as given on the type plate.
In case the claimed minimum working pressure $ \left(p_{\mathrm{min}}\right) $ of the meter is $ \geq 4 $ bar, the meter can be calibrated on one pressure. The working pressure range $ \left(p_{\mathrm{min}}\right. $ and $ p_{\mathrm{max}} $ ) is allowed to be within $ \frac{1}{2}\cdot p_{\mathrm{fix}} $ and $ 3\cdot p_{\mathrm{fix}}. $
The ultrasonic gas meter does not make use of an internal pressure or temperature sensor. The setting for pressure and temperature are a fixed value (fixed setting before calibration) in the parameter settings:
In the menu sequence $ p_{\mathrm{fix}} $ can be checked and represents the fixed pressure setting.
In the menu sequence $ T_{\mathrm{fix}} $ can be checked and represents the fixed temperature setting.
For a maximum working pressure $ \left(p_{\mathrm{m a x}}\right) $ of the ultrasonic gas meter above 60 bar(a), a maximum calibration pressure $ \left(p_{\mathrm{m a x}} \mathrm{c a l i b r a t i o n}\right) $ of around 60 bar(a) is allowed since this is the highest possible test pressure at any test location traceable to (inter)national standards.
Thermowell
A thermowell may be mounted at 2D-5D from the outlet of the meter.
Bi-directional flow measurement
During conformity assessment it is sufficient to verify a bi-directional meter in one direction only. For bi-directional flow measurement the outlet pipe and flow conditioner shall be identical to the inlet. The installation of a temperature sensor is at 2-5D from the outlet of the meter. For bi-directional applications an additional temperature sensor can be installed 2-5D upstream of the meter. For bi-directional applications the meter and pipe spools including the thermo well(s), shall be calibrated as a meter package during the examination for putting into use of the gas meter
OIML Certificate
OIML Member State The Netherlands
Number R137/2012-A-NL1-21.05
Project number 2518128
Alternative welded configuration of the gas meter
Page 6 of 6
The central meter body can be welded directly onto the flanges or to inlet and outlet pipes. The welding may not cause more than a 3% diameter step. The meter shall be installed as stated in "Installation requirements". The central meter body, including welded piping or welded flanges, shall be calibrated as a meter package during the examination for putting into use of the gas meter.
Maintenance
A transducer path pair and the main electronics board can be exchanged without deterioration of the metrological performance.
| Revision | Date | Description of the modification |
| Initial | 4 August 2021 | Initial issue |