R137/2012-A-NL1-21.05 Rev. 2

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-21.05 revision 2 Project number 3939292 Page 1 of 6

Issuing authority

NMi Certin B.V.

Person responsible: M.Ph.D. Schmidt

Applicant and Manufacturer

Endress+Hauser SICK GmbH+Co. KG Bergener Ring 27 01458 Ottendorf-Okrilla Germany

Identification of the certified type

An ultrasonic gas meter Type: FLOWSIC600-XT

Characteristics

See page 2 and further

This OIML Certificate is issued under scheme A

This Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML Type Evaluation Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

R 137-1:2012 "Gas meters"

Accuracy class 0,5

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation identified above. This Certificate does not bestow any form of legal international approval.

This certificate and supporting reports comply with the requirements of OIML-CS-PD-07 clause 6.2.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Type Evaluation Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 28 April 2025

Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

This document is digitally signed and sealed. The digital signature can be verified in the blue ribbon at the top of the electronic version of this certificate.

OIML Certificate

Number R137/2012-A-NL1-21.05 revision 2 Project number 3939292 Page 2 of 6

  • NMi-2518128-01 revision 1, dated 27 July 2023 that includes 65 pages.

The conformity was established by the results of tests and examinations provided in the associated reports:

  • NMi-3509954-01 dated 27 July 2023 that includes 33 pages.

Characteristics of the measuring instrument

In Table 1 the general characteristics of the measuring instrument are presented. In Table 4 gives an overview of the general characteristics of the family of instruments.

Table 1 General characteristics

Destined for the measurement ofGas volume
Environmental classesM2/E2
Accuracy class0.5
Minimum operating pressureatmospheric
Maximum operating pressuredepending on transducer type, see below
Ambient temperature range-40...+70℃
Process gas temperature range-40...+70℃
Designed forNon-condensing humidity&condensing humidity
OrientationHorizontal, vertical up and vertical down(all orientations)
Power supply voltage6-24VDC
Software identificationSee Table2
Path Angle60°for all meter types
TransducersTypeFrequency[kHz]pmax[bar]Beam angle[°]
153501058,9
2220520012,2
2620520011,7
2813520011,8
4620520015,9
S53501059,5
S620510311,9
S7135209,0
S813510310,6
K81351039,9
T21020510513,7
T21820510513,7
T22020520012,2
T22820520012,2
T2402052009,5
H2102051056,8
H2182051056,8

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-21.05 revision 2

Project number 3939292

Page 3 of 6

Table 2 Software Identification

Software versionsCRC ChecksumRemarks
1.05.000x78E1Initial release.
1.06.000x06F9-
2.01.000x1E05-

The software version and checksum can be accessed via the display of the gas meter.

Table 3 Meter types, path configuration & installation conditions

The meter consists of a cylindrical spool piece with 4 or 8 horizontal direct paths. Multiple configurations can be combined in a single housing according to the table below.

When the meter is equipped with a double meter configuration, the pay and check meter shall have a different indicator. A clear distinction between pay and check indicator shall be present, both meters must have their own nameplate with unique serial number.

FLOWSIC600-XT...Metrological relevant transducer pathsCheck paths(non-metrological relevant)Flow disturbanceMinimum installation conditionsup- and downstream[1]
Standard4-Mild2D-FC-8D-USM-3D
2plex41Mild2D-FC-8D-USM-3D
Quatro44Mild2D-FC-8D-USM-3D
Forte8-Mild&Severe5D-USM-3D

[1] xD = Minimum straight inlet piping length

FC = Flow conditioner: PTB design, FL600, drawing no. 9211779, d.d. 2014-05-19

USM = Ultrasonic gas meter

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-21.05 revision 2

Project number 3939292

Page 4 of 6

Table 4 General characteristics of the family of instruments

Standard(4Path),2plex(4Path),Quatro(4Path)和Forte(8Path)
Diameter NominalInternal boreMaximum $V_{\mathrm{max}}$ [m/s]Minimum
minimum [mm]maximum [mm]$V_{\mathrm{t}}$ [m/s]$V_{\mathrm{min}}$ [m/s]
3"(80mm)738539,222,450,31
4"(100mm)94,961102,54
6"(150mm)1421601,75
8"(200mm)1902101,57
10"(250mm)2352701,54
12"(300mm)27031537,841,5
14"(350mm)31534035,64
16"(400mm)36040035,48
18"(450mm)40545034,64
20"(500mm)450500
22"(550mm)495550
24"(600mm)540600
26"(650mm)58565033,96
28"(700mm)63070033,07
30"(750mm)67575032,08
32"(800mm)72080031,05
34"(850mm)76585029,34
36"(900mm)81090028,40
38"(950mm)85595027,49
40"(1000mm)900100027,09

Remarks regarding table 3:

[1] If higher values are chosen for $ Q_{\mathrm{min}} $ or $ Q_{t} $ and/or lower values for $ Q_{\mathrm{max}} $ it has to be taken into account that:

$$

  • \quad \mathrm {I f r a t i o} Q _ {\max } / Q _ {\min } \geq 5 0 \quad \mathrm {t h e n :} Q _ {\max } / Q _ {t} \geq 1 0 $$

$$

  • \quad \mathrm {I f r a t i o} Q _ {\max } / Q _ {\min } \geq 5 \mathrm {a n d} < 5 0 \quad \mathrm {t h e n :} Q _ {\max } / Q _ {\mathrm {t}} \geq 5 $$

[2] Depending on the internal bore, the flow rate in $ m^{3}/h $ can be calculated.

The corresponding flow rates can be calculated as follows:

$$ Q = v \cdot \frac {1}{4} \cdot \pi \cdot D ^ {2} \cdot 3 6 0 0 $$

Where:

$$ Q = \mathrm {f l o w r a t e} \quad [ \mathrm {m} ^ {3} / \mathrm {h} ] $$

$$ v = \mathrm {v e l o c i t y} \quad [ \mathrm {m} / \mathrm {s} ] $$

D = internal diameter [m]

[3] Example calculation for a 4" ultrasonic gas meter with an internal bore of 94,96 mm:

$$ Q _ {\max } = 3 9, 2 2 \cdot 1 / 4 \cdot 3, 1 4 1 5 9 \cdot 9 4, 9 6 ^ {2} \cdot 3 6 0 0 = 1 0 0 0 \mathrm {m} ^ {3} / \mathrm {h} $$

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-21.05 revision 2 Project number 3939292 Page 5 of 6

Installation conditions:

Installation requirements

Any components which could affect the gas flow must be avoided within the prescribed inlet pipe length. The inlet pipe must be designed as a straight pipe section of the same nominal diameter as the gas meter. The necessary straight pipe length is stated on the name plate of the meter.

The installation requirements, based on accuracy class and mild and/or severe flow disturbance are specified in table 2. The used flow conditioner (if applicable) is of type: "PTB design flow conditioner", type FL600, as documented in drawing number 9211779, dated 2014-05-19.

Working pressure

The spool piece and the transducers can be used up to the specified pressure as given in table 1.

The two pressures, at which the meter shall be calibrated, without intermediate adjustment over its full flow rate range during initial verification ( $ p_{\mathrm{min}} $ calibration and $ p_{\mathrm{max}} $ calibration),shall be equal to the operating pressure range ( $ p_{\mathrm{min}} $ and $ p_{\mathrm{max}} $ )as given on the type plate.

In case the claimed minimum working pressure $ \left(p_{\mathrm{min}}\right) $ of the meter is $ \geq 4 $ bar, the meter can be calibrated on one pressure. The working pressure range $ \left(p_{\mathrm{min}}\right. $ and $ p_{\mathrm{max}} $ ) is allowed to be within $ \frac{1}{2}\cdot p_{\mathrm{fix}} $ and $ 3\cdot p_{\mathrm{fix}}. $

The ultrasonic gas meter does not make use of an internal pressure or temperature sensor. The setting for pressure and temperature are a fixed value (fixed setting before calibration) in the parameter settings:

  • In the menu sequence $ p_{\mathrm{fix}} $ can be checked and represents the fixed pressure setting.

  • In the menu sequence $ T_{\mathrm{fix}} $ can be checked and represents the fixed temperature setting.

For a maximum working pressure ( $ p_{\mathrm{m a x}} $ )of the ultrasonic gas meter above 65 bar(a), a maximum calibration pressure ( $ p_{\mathrm{m a x}} $ calibration)of around 65 bar(a) is allowed since this is the highest possible test pressure at any test location traceable to (inter)national standards.

Thermowell

A thermowell may be mounted at 2D-5D from the outlet of the meter.

Bi-directional flow measurement

During conformity assessment it is sufficient to verify a bi-directional meter in one direction only. For bi-directional flow measurement the outlet pipe and flow conditioner shall be identical to the inlet. The installation of a temperature sensor is at 2-5D from the outlet of the meter. For bi-directional applications an additional temperature sensor can be installed 2-5D upstream of the meter. For bi-directional applications the meter and pipe spools including the thermowell(s), shall be calibrated as a meter package during the examination for putting into use of the gas meter.

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-21.05 revision 2

Project number 3939292

Page 6 of 6

Alternative welded configuration of the gas meter

The central meter body can be welded directly onto the flanges or to inlet and outlet pipes. The welding may not cause more than a 3% diameter step. The meter shall be installed as stated in "Installation requirements". The central meter body, including welded piping or welded flanges, shall be calibrated as a meter package during the examination for putting into use of the gas meter.

Maintenance

A transducer path pair and the main electronics board can be exchanged without deterioration of the metrological performance.

Revision history

This revision replaces the previous version.

RevisionDateDescription of the modification
Initial4 August 2021Initial issue
127 July 2023New Electronics, firmware, and transducers.
228 April 2025Manufacturer name change and software update. Previously named as“Sick Engineering GmbH”.