R137/2012-A-NL1-20.16 Rev. 1

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-20.16 revision 1 Project number 3594767 Page 1 of 6

Issuing authority

NMi Certin B.V.

Person responsible: M.Ph.D. Schmidt

Applicant and Manufacturer

KROHNE Altometer Kerkeplaat 12 3313 LC Dordrecht The Netherlands

Identification of the certified type

A ultrasonic gas meter

Types: ALTOSONIC V12

ALTOSONIC V12 Check

ALTOSONIC V12D

ALTOSONIC V12D Check

Characteristics

See page 2 and further

This OIML Certificate is issued under scheme A

This Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML Type Evaluation Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

R 137-1 (2012) "Gas meters"

Accuracy class

0,5 or 1,0 (depending on meter type, piping configuration and flow disturbance, see table 1 and 2 on the next page for a detailed overview)

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation identified above. This Certificate does not bestow any form of legal international approval.

This certificate and supporting reports comply with the requirements of OIML-CS-PD-07 clause 6.2.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Type Evaluation Reports is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 20 December 2022

Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

Reproduction of the complete document only is permitted.

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-20.16 revision 1

Project number 3594767

Page 2 of 6

The conformity was established by the results of tests and examinations provided in the associated reports:

  • No. CVN-710381-01 Rev. 1a dated 25 March 2009 that includes 45 pages;

  • No. NMi 1902254-01 dated 21 December 2018 that includes 33 pages;

  • No. NMi-14200114-01 dated 29 October 2014 that includes 18 pages;

  • No. NMi-1901377-01 dated 8 October 2019 that includes 22 pages;

  • No. NMi-2408281-01 dated 31 January 2020 that includes 17 pages.

  • No. NMi-2533860-01 dated 24 December 2020 that includes 12 pages.

Characteristics of the measuring instrument

The general characteristics of the family of instruments are presented in table 1 (ALTOSONIC V12), table 2 (ALTOSONIC V12D) and table 3 (both meter types). Table 4 presents the operating ranges per transducer type (both meter types). The flow characteristics are presented in table 5 (ALTOSONIC V12) and table 6 (ALTOSONIC V12D).

Both meter types, ALTOSONIC V12 and ALTOSONIC V12D, can also be equipped with an additional non-metrological relevant index, meter types: ALTOSONIC V12 Check and ALTOSONIC V12D Check.

Table 1 General characteristics ALTOSONIC V12 and ALTOSONIC V12 Check

Characteristics
Path configurationALTOSONIC V12Metrological relevant5 horizontal reflecting measuring paths
Optional, non-metrological relevantOptional,1 vertical reflecting measuring path
ALTOSONIC V12 CheckMetrological relevant5 horizontal reflecting measuring paths
Non-metrological relevant1 vertical reflecting measuring path
Accuracy classALTOSONIC V12,ALTOSONIC V12 CheckMild flow disturbances10D-USM0.5
2D-FC-3D-USM0.5
Severe flow disturbances2D-FC-3D-USM0.5
10D-USM1.0

Table 2 General characteristics ALTOSONIC V12D and ALTOSONIC V12D Check

Characteristics
Path configurationALTOSONIC V12DMetrological relevant5 horizontal direct measuring paths
Optional, non-metrological relevantOptional,1 vertical reflecting measuring path
ALTOSONIC V12D CheckMetrological relevant5 horizontal direct measuring paths
Non-metrological relevant1 vertical reflecting measuring path
Accuracy classALTOSONIC V12D,ALTOSONIC V12D CheckMild flow disturbances2D-FC-8D-USM1,0
Severe flow disturbances2D-FC-8D-USM1,0

Remarks regarding table 1 and 2: see next page.

OIML Certificate

Number R137/2012-A-NL1-20.16 revision 1

Project number 3594767

Page 3 of 6

Remarks regarding table 1 and 2:

  • xxD = Minimal length of straight inlet/outlet piping (where x can be 0..9).

  • FC = Flow conditioner.

Hereinafter:

  • USM = Ultrasonic gas meter.

  • the ALTOSONIC V12 & ALTOSONIC V12 Check are also referred to as ALTOSONIC V12 (Check);

  • the ALTOSONIC V12D & ALTOSONIC V12D Check are also referred to as ALTOSONIC V12D (Check).

Table 3 General characteristics ALTOSONIC V12 (Check) and ALTOSONIC V12D (Check)

Meter typeALTOSONIC V12(Check)和ALTOSONIC V12D(Check)
Maximum pressureSee table4
Ambient temperature range-40℃/+55℃
Gas temperature rangeSee table4
Intended for the measurement ofNatural gas
Designed forCondensing humidity
OrientationHorizontal, vertical up and vertical down(all orientations)
Flow directionBi-directional
Path angle60°...80°
Power supply voltage24VDC
Software identification:-Version number-ChecksumKAFKA2.0.0.0D4F94254KAFKA2.0.0.1F0591D45KAFKA2.0.0.20EE46EDAKAFKA2.0.0.30x3834247E
KAFKA2.0.0.50xCF968C03

Table 4 Operating ranges of the instrument per used transducer type

The gas temperature range and maximum operating pressure of the instrument is limited to the ranges as indicated below based on transducer type.

Transducer type[2]ALTOSONIC V12(Check)andALTOSONIC V12D(Check)Gas temperature rangeTmin...Tmax[℃]Maximum pressurePmax[bar(g)]Sound frequency[kHz]
G5.00-20...+70[1]150270
G5.01431
G5.02204
G5.03431
G6.00-20...+100[1]150
G6.01431
G6.02204
G6.03431
G7.01-20...+100[1]150330
G7.04100150

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-20.16 revision 1

Project number 3594767

Page 4 of 6

Remarks regarding table 4:

[1] The maximum gas temperature also depends on the chosen $ Q_{\mathrm{min}} $ as specified in table 5 or 6. The resulting limiting value for $ T_{\mathrm{max}} $ is normative.

[2] For severe flow disturbances the transducers are flush mounted (no protrusion inside the meter housing).

Table 5 General characteristics of the family of instruments - ALTOSONIC V12 (Check)

ALTOSONIC V12(Check)
Diameter NominalMinimum $Q_{\mathrm{min}}$Minimum $Q_{\mathrm{min}}$Maximum $Q_{\mathrm{max}}$Maximum $Q_{\mathrm{max}}$Maximum $Q_{\mathrm{max}}$Minimum $Q_{\mathrm{t}}$$P_{\mathrm{min}}^{[4]}$$P_{\mathrm{max}}$
$T_{\mathrm{gas}}\leq50^{\circ}\mathrm{C}$$T_{\mathrm{gas}}\leq100^{\circ}\mathrm{C}$Mild & Severe disturbanceG5.00...G6.03transducers[m3/h]Mild disturbanceG7.01 andG7.04transducers[m3/h]Severe disturbanceG7.01 andG7.04transducers[m3/h]$Q_{\mathrm{t}}$[m3/h]$P_{\mathrm{min}}^{[4]}$[barg]$P_{\mathrm{max}}$[barg]
[m3/h][m3/h]7751.010910391,0According to table 4
4"(100mm)10127751.010910391,0
6"(150mm)16301.8002.2802.050901,5
8"(200mm)20453.1004.0003.6001552,0
10"(250mm)25804.3006.0005.4002152,5
12"(300mm)351206.4008.2007.3803203,0
14"(350mm)451508.10010.5009.4504053,5
16"(400mm)6019510.10013.00011.7005054,0
18"(450mm)10024012.40016.00014.4006204,5
20"(500mm)15028514.70019.00017.1007355,0
24"(600mm)18037519.40025.00022.5009706,0
26"(650mm)25050021.70028.00025.20010857,0
28"(700mm)30060024.80032.00028.80012407,0
30"(750mm)40080027.10035.00031.50013557,0

Remarks regarding table 5:

[1] If higher values are chosen for $ Q_{\mathrm{min}} $ or $ Q_{t} $ and/or lower values for $ Q_{\mathrm{max}} $ , it has to be taken into account that:

$$

  • \quad \text {I f r a t i o} Q _ {\max } / Q _ {\min } \geq 5 0 \quad \text {t h e n}: Q _ {\max } / Q _ {\mathrm {t}} \geq 1 0 $$

$$ \text {I f r a t i o} Q _ {\max } / Q _ {\min } \geq 5 \text {a n d} < 5 0 \quad \text {t h e n}: Q _ {\max } / Q _ {\mathrm {t}} \geq 5 $$

[2] The maximum $ Q_{\mathrm{m a x}} $ depends on the transducer type and installation conditions (mild or severe flow disturbances).

[3] The maximum operating pressure ( $ P_{\mathrm{m a x}} $ )depends on the installed transducer according to table 4.

[4] Regarding $ p_{\min} $ and $ p_{\max} $ see also the section "Installation conditions, working pressure" on page 6.

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-20.16 revision 1

Project number 3594767

Page 5 of 6

Table 6 General characteristics of the family of instruments - ALTOSONIC V12D (Check)

ALTOSONIC V12D(Check)
Diameter NominalMinimum $Q_{\mathrm{min}}$Minimum $Q_{\mathrm{min}}$Maximum $Q_{\mathrm{max}}$Maximum $Q_{\mathrm{max}}$Maximum $Q_{\mathrm{max}}$Minimum $Q_{\mathrm{t}}$$P_{\mathrm{min}}^{[4]}$$P_{\mathrm{max}}$
$T_{\mathrm{gas}}\leq50^{\circ}\mathrm{C}$$T_{\mathrm{gas}}\leq100^{\circ}\mathrm{C}$Mild & Severe disturbanceG5.00...G6.03transducers[m3/h]Mild disturbanceG7.01/G7.04transducers[m3/h]Severe disturbanceG7.01/G7.04transducers[m3/h]$Q_{\mathrm{t}}$[m3/h]$P_{\mathrm{min}}^{[4]}$[barg]$P_{\mathrm{max}}$[barg]
4"(100mm)10127751.010950391,0According to table 4
6"(150mm)16301.8002.2802.130901,5
8"(200mm)20453.1004.0003.7501552,0
10"(250mm)25804.3006.0005.6502152,5
12"(300mm)351206.4008.2007.7003203,0
14"(350mm)451508.10010.5009.8504053,5
16"(400mm)6019510.10013.00012.2005054,0
18"(450mm)10024012.40016.00015.0006204,5
20"(500mm)15028514.70019.00017.8007355,0
24"(600mm)18037519.40025.00023.4509706,0
26"(650mm)25050021.70028.00026.25010857,0
28"(700mm)30060024.80032.00030.00012407,0
30"(750mm)40080027.10035.00032.80013557,0

Remarks regarding table 6:

[1] If higher values are chosen for $ Q_{\min} $ or $ Q_{t} $ and/or lower values for $ Q_{\max} $ it has to be taken into account that:

$$

  • \quad \text {I f r a t i o} Q _ {\max } / Q _ {\min } \geq 5 0 \quad \text {t h e n}: Q _ {\max } / Q _ {t} \geq 1 0 $$

$$

  • \quad \text {If r a t i o} Q _ {\max } / Q _ {\min } \geq 5 \text {a n d} < 5 0 \quad \text {t h e n :} Q _ {\max } / Q _ {\mathrm {t}} \geq 5 $$

[2] The maximum $ Q_{\mathrm{m a x}} $ depends on the transducer type and installation conditions (mild or severe flow disturbances).

[3] The maximum operating pressure ( $ P_{\mathrm{m a x}} $ )depends on the installed transducer according to table 4.

[4] Regarding $ p_{\mathrm{min}} $ and $ p_{\mathrm{max}} $ see also the section "Installation conditions, working pressure" on page 6.

OIML Certificate

OIML Member State The Netherlands

Number R137/2012-A-NL1-20.16 revision 1

Project number 3594767

Page 6 of 6

Installation conditions:

Installation requirements

Any components which could affect the gas flow must be avoided within the prescribed inlet pipe length. The inlet pipe must be designed as a straight pipe section of the same nominal diameter as the gas meter. The necessary straight pipe length is stated on the name plate of the meter.

The installation requirements, based on accuracy class and mild and/or severe flow disturbance are specified in table 1 and 2. The used flow conditioner (if applicable) is of type: "KROHNE flow conditioner".

Working pressure

The spool piece and the transducers can be used up to the specified pressure in table 4 ( $ p_{\mathrm{max}} $ depending on transducer type).

The two pressures, at which the meter shall be calibrated, without intermediate adjustment over its full flow rate range during initial verification ( $ p_{\mathrm{min}} $ calibration and $ p_{\mathrm{max}} $ calibration),shall be equal to the operating pressure range ( $ p_{\mathrm{min}} $ and $ p_{\mathrm{max}} $ )as given on the type plate.

In case the claimed $ P_{\mathrm{min}} $ of the meter is $ \geq10 $ bar, and $ P_{\mathrm{max}}-P_{\mathrm{min}}\leq 30 $ bar, only 1 calibration pressure ( $ p_{\mathrm{calibration}} $ )is necessary.

For a maximum working pressure ( $ p_{\mathrm{m a x}} $ )of the ultrasonic gas meter above 60 bar(a), a maximum calibration pressure ( $ p_{\mathrm{m a x}} $ )of around 60 bar(a) is allowed since this is the highest possible test pressure at any test location traceable to (inter)national standards.

Meter in series

It is allowed to install the ALTOSONIC V12 (Check) and ALTOSONIC V12D (Check) meters in series with another ultrasonic meter by directly connecting the outlet of the first meter to the inlet of the second meter, while applying the following conditions:

  • Both meters are calibrated at the same time, being mounted together;

  • The installation conditions of both meter types are fulfilled.

Reduced bore

In case the ALTOSONIC V12 (Check) or ALTOSONIC V12D (Check) is applied in series with a reduced bore meter, the following condition is also applicable:

  • If the ALTOSONIC V12 (Check) or ALTOSONIC V12D (Check) is mounted on the second position, then the angle of inclination of the reduced bore shall be less than $ 7^{\circ} $

Certificate history:

This revision replaces the previous version.

RevisionDateDescription of the modification
024 December 2020Initial
120 December 2022Updated the software to V2.0.05