R137/2012-NL1-2017.03

Applicant
Elster GmbH
Issuer
NL1
Issuing year
2017
Status
Valid

Source PDF (certificates/R137/2012/r137-2012-nl1-2017-03.pdf) · extracted by OCR from the digitalized tier of the corpus.

R137/2012-NL1-2017.03

OIML Certificate of Conformity

OIML Member State The Netherlands

Number R137/2012-NL1-17.03 Project number 16200535 Page 1 of 4

Issuing authority NMi Certin B.V.

Person responsible: C. Oosterman

Applicant and Elster GmbH

Manufacturer Steinern Straße 19-21

55252 Mainz-Kastel

Germany

Identification of the An Ultrasonic Gas Meter certified type Type: Q.Sonic $ ^{m a x} $

Characteristics See page 2 and further

This Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML Type Evaluation Report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

R 137-1 (2012) "Gas meters"

Accuracy class 0,5

This Certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation identified above. This Certificate does not bestow any form of legal international approval.

Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate was issued, partial quotation of the Certificate and of the associated OIML Type Evaluation Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1

C. Oosterman Head Certification Board

NMi Certin B.V.

Hugo de Grootplein 1

3314 EG Dordrecht

the Netherlands

T +31 78 6332332

certin@nmi.nl

www.nmi.nl

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

The notification of NMi Certin B.V. as Issuing Authority can be verified at www.oiml.org

OIML Certificate of Conformity

OIML Member State

Number R137/2012-NL1-17.03

The Netherlands

Project number 16200535

Page 2 of 4

The conformity was established by the results of tests and examinations provided in the associated report(s):

  • No. NMi-9200288-01 dated 18 January 2012 that includes in total 48 pages;

  • No. NMi-9200288-03 dated 3 July 2012 that includes in total 47 pages;

  • No. NMi-16200535-01 revision 1 dated 1 September 2017 that includes 35 pages.

Characteristics of the measuring instrument

In Table 1 the general characteristics of the measuring instrument are presented.

Table 2 gives an overview of the general characteristics of the family of instruments.

The construction of the measuring instrument is recorded in the Documentation folder no. T10586-1.

Table 1 General characteristics

Destined for the measurement ofGas volume
Environmental classesM1/E2
Accuracy class0.5
Working pressureThe meter is either programmed with a density and viscosity setting, corresponding to the applied gas at preset pressure or using live pressure value from the optional pressure sensor. When using a preset value the maximum operating range for pressure $ p_{\mathrm{max}}/p_{\mathrm{min}} $ is 6,25 symmetrically divided around the preset pressure.
Gas temperature range-40-+100℃ gas meter-30-+80℃ gas meter with EVCD
Ambient temperature range(condensing)-40-+70℃ gas meter-40-+55℃ gas meter with EVCD
Designed forNon condensing humidity/condensing humidity
OrientationHorizontal/vertical up/vertical down/all orientations
Power supply voltage18-30V DC
Software identificationVersionChecksum
Basic System02.10.01.00037C01880E
Archive02-07-A420652AC
Postprocessing02-08-C459B14C1
Density & Viscosity02.04.00.0001BB87257F
Geometry Correction02.04.00.0001D8D14A5F
Modbus02-20-BF4348ED8
USM02.10.00.00003DE0303F
NGQFB01.00.07B7568107
NGQMB01.00.100C432F97
Gas Quality02-09-CC83315E2
Flow Conversion02-07-D587979DE

OIML Certificate of Conformity

OIML Member State The Netherlands

Number R137/2012-NL1-17.03

Project number 16200535

Page 3 of 4

Table 2 General characteristics of the family of instruments

Diameter sizeMinimum $ Q_{\mathrm{min}} $ [m3/h]Minimum $ Q_{\mathrm{t}} $ [m3/h]Maximum $ Q_{\mathrm{max}} $ [m3/h]
InchDNInternal bore [mm]
410097131001000
6150146182202200
139162002000
8200190304003500
180273503200
10250240485905700
230445405200
12300295738608300
280667807500
1435032585100010000
305759008800
20500488,9200210021000
431,8160160016000
24600590,9295300030000
532,22240240024000

OIML Certificate of Conformity

OIML Member State

Number R137/2012-NL1-17.03

The Netherlands

Project number 16200535

Page 4 of 4

Conditions for conformity assessment:

Installation of the ultrasonic gas meter

For mild and severe flow disturbances the meter needs to be installed with the following minimum piping configuration as upstream inlet:

  • mild disturbances: 5D straight piping.

  • severe disturbances: 5D straight piping + CPA50 type A flow conditioner + 5D straight piping.

The outlet piping shall comply with:

  • 3D straight piping (for both mild and severe disturbances).

Any components which could affect the gas flow must be avoided within the above prescribed inlet pipe length. The necessary straight pipe length is stated on the name plate of the meter. The inlet pipe must be designed as a straight pipe section of the same nominal diameter as the gas meter with a maximum tolerance of +/- 3%.

Bi-directional flow measurement

During conformity assessment it is sufficient to verify a bi-directional meter only in one direction.

Temperature sensor

The installation of a temperature sensor is at 2-5D from the outlet of the meter. For bi-directional applications an additional temperature sensor can be installed 2-5D upstream of the meter. The pipe spools including the thermo well(s) shall be installed and considered during the examination for putting into use of the gas meter.

Alternative welded installation

The meters can be welded directly without flanges to the inlet and outlet pipes. The inlet and outlet pipe length shall be according the requirements as prescribed above for mild and/or severe disturbances. The complete meter package (meter including welded piping) shall be calibrated in order to guarantee compliance with Class 0.5.