Skip to content

← R137

R137/2012-NL1-2014.03 Rev. 2

Applicant
Flow Meter Group B.V.
Issuer
NL1
Issuing year
2016
Status
Valid

Source PDF (certificates/R137/2012/r137-2012-nl1-2014-03-rev2.pdf) · extracted by OCR from the digitalized tier of the corpus.

R137/2012-NL1-2014.03 Rev. 2

OIML Certificate of Conformity

OIML Member State The Netherlands

Number R137/2012-NL1-14.03 revision 2 Project number SO15202424 Page 1 of 5

Issuing authorityNMi Certin B.V. Person responsible: C. Oosterman
Applicant and manufacturerFlow Meter Group B.V. Meniststraat 5c 7091 ZZ Dinxperlo The Netherlands
Identification of the certified typeA turbine gas meter Type: FMT-L, FMT-S, FMT-Lx and FMT-Dc
CharacteristicsSee page 2 up to and including page 5.

This certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML Type Evaluation Reports) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML):

R 137-1 (2012) "Gas meters"

Accuracy class

This certificate relates only to the metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML International Recommendation identified above. This certificate does not bestow any form of legal international approval.

Important note: Apart from the mention of the certificate's reference number and the name of the OIML Member State in which the certificate was issued, partial quotation of the certificate and of the associated OIML Type Evaluation Report(s) is not permitted, although either may be reproduced in full.

Issuing Authority

NMi Certin B.V., OIML Issuing Authority NL1 13 October 2016

C. Oosterman Head Certification Board

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability.

The notification of NMi Certin B.V. as Issuing Authority can be verified at www.oiml.org

OIML Certificate of Conformity

OIML Member State The Netherlands

Number R137/2012-NL1-14.03 revision 2 Project number SO15202424 Page 2 of 5

The conformity was established by the results of tests and examinations provided in the associated (OIML Type Evaluation) Reports:

  • No. NMi-14200104-01 dated 15 September 2014 that includes 14 pages;

  • No. NMi-14200702-01 dated 18 December 2014 that includes 13 pages;

  • No. NMi-14200712-01 dated 23 December 2014 that includes 35 pages;

  • No. VSL-3255499-01 dated 28 January 2015 that includes 6 pages.

Characteristics of the gas meter:

Table 1 gives the general characteristics of all three turbine meter types. Table 2 and 3 on the following pages specify in detail the characteristics and essential parts of the turbine meters. The construction of the measuring instrument is recorded in the Documentation folder no. T10466-6.

Destined for the measurement ofGas volume
Mechanical classM2
Electromagnetic classNot applicable (the meter has no electronics)
Ambient temperature range-40℃/+70℃
Gas temperature range-40℃/+70℃
OrientationHorizontal/Vertical up/Vertical down(all orientations)
Flow directionUni-directional(indicated with arrow)
OIML Member State The Netherlands

OIML Certificate of Conformity

Number R137/2012-NL1-14.03 revision 2

Project number SO15202424

Page 3 of 5

Table1
Main characteristics1.2Essential characteristics1.4Conditional parts1.1Essential parts
TypeDN bodyDN cartr.G-valueQmaxQtQminNumber of drumsBefore and after commaControl elementBlade angleBlade qtyBearing dimensionsInside diameterxOutside diameterxWidth in mm
MR1:200..101bar(g)MR1:304..101bar(g)MR1:508..101bar(g)standard bearings0..101bar(g)Optional bearings4...101bar(g)
standard bearings0..101bar(g)Optional bearings4...101bar(g)
-mmmm-m3/hm3/hm3/hm3/hm3/h-m3/hdegree-frontrearfrontrear
FMT-Lx orFMT-Dc5080100160328537/10,0255123x6x32x5x23x8x43x6x3
1602505012,5854514
250400802012,583516
FMT-L,FMT-S,FMT-Lx orFMT-Dc8080100160328537/10,0255123x6x32x5x23x8x43x6x3
1602505012,5854514
250400802012,583516
FMT-Lx orFMT-Dc801001602505012,5857/10,0255123x8x43x6x35x10x33x6x3
250400802012,584514
40065013032,52012,53516
FMT-L,FMT-S,FMT-Lx orFMT-Dc1001001602505012,5857/10,0255123x8x43x6x35x10x33x6x3
250400802012,584514
40065013032,52012,53516
FMT-Lx orFMT-Dc10015040065013032,52012,57/10,0255125x11x55x11x55x16x55x11x5
65010002005032204514
100016003208050328/00,23516
FMT-L,FMT-S,FMT-Lx orFMT-Dc15015040065013032,52012,57/10,0255125x11x55x11x55x16x55x11x5
65010002005032204514
100016003208050328/00,23516
FMT-Lx orFMT-Dc15020065010002005032207/10,0255125x16x55x16x58x22x75x16x5
100016003208050328/00,24514
1600250050012580508/00,23516
FMT-Lx orFMT-Dc25020065010002005032207/10,0255125x16x55x16x58x22x75x16x5
100016003208050328/00,24514
1600250050012580508/00,23516
OIML Member State The Netherlands

OIML Certificate of Conformity

Number R137/2012-NL1-14.03 revision 2

Project number SO15202424

Page 4 of 5

Table2
Main characteristics1.2Essential characteristics1.4Conditional parts1.1Essential parts
TypeDN bodyDN cartr.G-valueQmaxQtQminNumber of drumsBefore and after commaControl elementBlade angleBlade qtyBearing dimensionsInside diameterxOutside diameterxWidthin mm
0..101bar(g)MR 1:304..101bar(g)MR 1:500..101bar(g)
standard bearingsOptional bearings
-mmmm-m3/hm3/hm3/hm3/h-m3/hdegree-frontrearfrontrear
FMT-Lx orFMT-Dc2002501000160032050328/00,2452410x26x86x16x512x28x86x19x6
16002500500805045/3024
25004000800125803024
FMT-S,FMT-Lx orFMT-Dc2502501000160032050328/00,2452410x26x86x16x512x28x86x19x6
16002500500805045/3024
25004000800125803024
25004000800125803024
FMT-Lx orFMT-Dc2503001600250050080508/00,2452412x28x86x19x617x40x128x22x7
250040008001258045/3024
4000650013002161303024
FMT-S,FMT-Lx orFMT-Dc3004001600250050080508/00,2452412x28x86x19x617x40x128x22x7
250040008001258045/3024
4000650013002161303024
4000650013002161303024
FMT-Lx orFMT-Dc4003001600250050080508/00,2452415x35x118x22x717x40x128x22x7
250040008001258045/3024
4000650013002161303024
FMT-S,FMT-Lx orFMT-Dc4004002500400080133808/00,2452415x35x118x22x717x40x128x22x7
40006500130021613045/3024
4000650013002161303024
FMT-Lx orFMT-Dc4005004000650013002161308/00,2452420x47x1410x26x820x47x2010x26x8
650010000200033320045/3024
100001600032005333203024
FMT-Lx orFMT-Dc50060065001000020003332008/00,2452420x47x2010x26x820x47x2010x26x8
1000016000320053332045/3024
160002500050008004003024
FMT-Lx orFMT-Dc5006004000650013002161308/00,2452420x47x2010x26x820x47x2010x26x8
650010000200033320045/3024
100001600032005333203024

OIML Certificate of Conformity

OIML Member State The Netherlands

Number R137/2012-NL1-14.03 revision 2 Project number SO15202424 Page 5 of 5

Installation conditions:

The FMT-L meter can be equipped with two different flow straighteners. When the large one is installed a straight inlet pipe of 1xDN is applicable. With the small one installed a straight inlet pipe of 5xDN shall be applied.

The FMT-S can only be equipped with the large flow straightener, and therefore needs a minimum of 1xDN straight pipe length in front of the inlet.

The FMT-Lx meter has no installation requirements concerning straight inlet tubing. The inlet flow straightener is identical to the FMT-S configuration due to the fact that the FMT-Lx and FMT-S have identical cartridges.

The FMT-Dc meter consists of two FMT-Lx cartridges in one housing. The same installation conditions and flow straighteners are applicable for the FMT-Dc as given for the FMT-Lx.

The necessary straight pipe length is stated on the name plate of the meter.

Any components which could affect the gas flow must be avoided within the prescribed inlet pipe length. The inlet pipe must be designed as a straight pipe section of the same nominal diameter as the gas meter.

Certificate history:

This revision replaces the previous versions.

RevisionDateDescription of the modification
Initial24 December 2014-
130 July 2015FMT-Dc meter type added to document
213 October 2016Editorial change