R117/2019-A-SE1-22.01
| OIML Member State Sweden | OIML Certificate No. R117/2019-A-SE1-22.01 |
| OIML CERTIFICATE ISSUED UNDER SCHEME A | |
| OIML Issuing Authority Name: RISE Research Institutes of Sweden AB Address: Box 857, SE-50115 Borås, Sweden Person responsible: Martin Tillander | |
| Applicant Name: Dover Fueling Solutions UK Limited Filial (DFS) Postal address: PO.Box 50559, SE-205 15 Malmö Visiting address: Hanögatan 8, SE-211 24 Malmö | |
| Manufacturer Name: Dover Fueling Solutions UK Limited Address: Unit 3, Baker Road, West Pitkerro Industrial Estate, Dundee DD5 3RT, United Kingdom Name: Wayne Industria e Comercio Ltda Address: Estrada Timbó 126--Higienópolis, 2106-280—Rio de Janeiro, --RJ—Brazil Name: Tokheim India Pvt. Ltd Address: Building No. 2, Plot No. 66, TTC Industrial Area, MIDC Mahape, Navi Mumbai – 400710 Maharashtra, India | |
| Identification of the certified type (the detailed characteristics will be defined in the additional pages and in appendix to this certificate) Wayne Helix XXXX Fuel Dispenser (there XXXX stands for a number between 1000 and 6000) | |
| Designation of the module (if applicable) ... | |
| This OIML Certificate attests the conformity of the above identified type (represented by the sample(s) identified in the OIML type evaluation report) with the requirements of the following Recommendation of the International Organization of Legal Metrology (OIML): OIML R117 Edition (year): 2019 For accuracy class 0,5 or 1,0 (for LPG) | |
| OIML Certificate No. R117/2019-A-SE1-22.01 | ||
| This OIML Certificate relates only to metrological and technical characteristics of the type of measuring instrument covered by the relevant OIML Recommendation identified above. This OIML Certificate does not bestow any form of legal international approval. The validity of this certificate can be verified by RISE. | ||
| The conformity was established by the results of tests and examinations provided in the associated OIML type evaluation report: No.1122498... dated 2022-09-30 that includes 17 pages | ||
| The technical documentation relating to the identified type is contained in documentation file which is included in OIML type evaluation report mentioned above. OIML Certificate History | ||
| Revision No. | Date | Description of the modification |
| 0 | 2022-09-30 | First edition |
| Identification, signature and stamp The OIML Issuing Authority RISE Research Institutes of Sweden AB Date 2022-09-30 | ||
| Important note: Apart from the mention of the Certificate's reference number and the name of the OIML Member State in which the Certificate is issued, partial quotation of the Certificate and of the associated OIML type evaluation report(s) is not permitted, although either may be reproduced in full. | ||
ANNEX to an OIML Certificate of Conformity
In respect of (type of instrument)
A family of Wayne Helix 1000 Fuel dispenser, Wayne Helix 2000 Fuel dispenser, Wayne Helix 4000 Fuel dispenser, Wayne Helix 5000 Fuel dispenser, Wayne Helix 6000 Fuel dispenser, for petrol, kerosene, diesel, ethanol, FAME/RME, HVO fuel, DEF (Diesel Exhaust Fluid), Windshield liquid or LPG, including ATC conversion device (optional).
Product names (see pictures 1-9):
Wayne Helix 1000 Fuel dispenser
Wayne Helix 2000 Fuel dispenser
Wayne Helix 4000 Fuel dispenser
Wayne Helix 5000 Fuel dispenser
Wayne Helix 6000 Fuel dispenser
Product designation:
A and B side is recognized by removing one of the doors from the hydraulic cabinet and look at the sheet metal support for pumping units (suction-dispensers) or meters (remote-dispensers). There is the letter A or B punched in to the sheet metal as can be seen in the pictures below.
remote pump
Measuring system (fuel dispenser) description
A complete measuring system consists of one electronic module and one to six hydraulic modules. The hydraulic modules are placed in one or two cabinets (B2B). One electronic module can serve both cabinets, for example a combination of LPG and petrol. If one pump and air separator is serving more than one nozzle simultaneously the total maximum flowrate through these nozzles is limited by the air separator (90 l/min per air separator) and the volume sensor (according to "Volume sensor flowrate
range under "Measurement range"). For higher flowrate another hydraulic module have to serve the same nozzle. For further information see block diagram under "Hydraulic function modules".
Electronic module function
(iGEM) is an electronic subsystem and it consists mainly of: calculator, indicating device, and keyboard with preset. This module can handle up to 6 motors, 6 pulse transmitters (each handling a single equipped or a duplex volume sensor), 12 nozzles, 14 solenoid valves and in case of ATC one temperature compensation module. The electronic module is able to serve up to four customers at a time (called 4 User).
Hydraulic function modules
Measurement transducer function is a hydraulic subsystem and it consists mainly of: volume sensor (single equipped side A, single equipped side B or duplex) and pulse transmitter. In case of ATC automatic temperature compensation) each volume sensor is equipped with a temperature sensor, all connected to one temperature compensation module.
Pump and air separator function is a hydraulic subsystem and it consists mainly of: Compact Pumping Unit (CPU) including air separator, motor and non-return valve and for LPG a differential valve.
Regulation function is a hydraulic subsystem and it consists mainly of: solenoid valves for flow rate regulation, blending and on/off.
Delivery function is a hydraulic subsystem and it consists mainly of: hoses, nozzles and nozzle switch, this includes also satellite function.
Central pump function is a hydraulic subsystem and it consists mainly of: delivery of air free liquid according to OIML R117-1, item 5.1.3 and security valve.
If the measuring system is equipped with a central pump (an additional "R") an external (central) pump is used instead of an internal pump. The external system must comply with OIML 117-1, item 5.1.3 (i.e. it shall be equipped with an arrangement that prevents air to come into the system). The same applies to additive injection.
During verification it shall be possible to verify each volume sensor separately.
Block diagram describing permitted ways, and some restrictions, to produce one delivery product for one customer. Each such configuration shall be a subset of the diagram. To produce multiple delivery products for a customer the block diagram is used several times.
Picture 1: Wayne Helix 5000 Fuel dispenser
Picture 2: Wayne Helix 5000 Fuel dispenser with Wayne iXPay $ ^{TM} $ secure payment platform.
Picture 3: Wayne Helix 2000 Fuel dispenser Picture 4: Wayne Helix 1000 Fuel dispenser
Picture 5: Wayne Helix 6000 Fuel dispenser
Picture 6: Wayne Helix 6000 Fuel dispenser with Wayne iXPay $ ^{TM} $ secure payment platform.
Picture 7: Wayne Helix 6000 Fuel dispenser Picture 8: Wayne Helix 1000 Fuel dispenser B2B LPG
Picture 9: Wayne Helix 6000 Fuel dispenser Metallic head
Components included
Wayne Helix Fuel dispenser families may include the following components.
Measurement transducer function
Volume sensor (petrol, petrol/ethanol mixtures, kerosene diesel or diesel/FAME/RME/HVO mixtures)
Wayne iMeter Duplex (DM2-2) or
Wayne iMeter Single equipped side A (DM2-1) or Wayne iMeter Single equipped side B (DM2-1) or Wayne iMeter2 Duplex (DM2-2) or Wayne iMeter2 Single equipped side A (DM2-1) or Wayne iMeter2 Single equipped side B (DM2-1) or Wayne $ \mathrm{X f l o}^{\mathrm{T M}} $ Duplex (magenta or silver id label) or Wayne $ \mathrm{X f l o}^{\mathrm{T M}} $ Single sided A (yellow id label) or Wayne $ \mathrm{X f l o}^{\mathrm{T M}} $ Single sided B (cyan id label)
Volume sensor (DEF) Volume sensor (DEF and Windshield liquid)
Wayne iMeter2 DM2-X DEF Wayne iMeter DM2-X UREA
Volume sensor (LPG) (including differential valve)
Petrolmeccanica LPG6000 MA (with mechanical adjustment)
Petrolmeccanica LPG6000 WA (without mechanical adjustment)
Tokheim GPL700
Pulse transmitter
| Wayne PNo | i-Meter | i-Meter2 | XfloTM | LPG |
| WM001682-0001 | NFC | NFC | NFC | |
| WM001682-0005 | NFC | NFC | NFC | |
| WM001682-0007 | NFC | NFC | ||
| WM001682-0009 | FC | |||
| WM001682-0011 | FC | |||
| WM011529-0001 | FCI | |||
| WM019142-0001 | FCI | |||
| WM019142-0003 | FCI |
FC=Flow Compensating, FCI = Flow Compensating Individual(XWIP), NFC= Not Flow Compensating (WIP)
Board temperature module, phase 2, Wayne WM044193-000X
0-1 pc****
Software Board temperature module, W&M checksum EB6A, or 7EA9
0-1 pc****
Temperature sensor, 0-10 pcs****
Wayne WM040341-0001
Pump and air separator function, not LPG
Compact Pumping Unit (CPU) $ ^{*} $
| Wayne | i-Meter | i-Meter2 | XfloTM |
| WM002219-0001 | D,H | D,H | |
| WM002219-0002 | P,K,E | P,K,E | |
| WM018719-0001 | P,K,D,H,E,F | P,K,D,H,E,F | P,K,D,H,E,F |
| WM016720-0001 | P,K,D,H,E,40l/min | ||
| WM016720-0003 | P,K,D,H,70l/min |
P = petrol, K = kerosene, D = diesel, E = ethanol, F=FAME/RME, H=HVO
Anti-Foaming pipe 0-1 pcs
DFS WR002578
(Can be installed in any above mentioned
pumping units)
Cover
(between CPU and meter) **
Wayne 167360 or Wayne WM000635 or Wayne WM009207 or Wayne WM011470 or Wayne WM011337 or Wayne WM018512-000X or Wayne WM011485-000X
Motor **, one per CPU
0,75 kW, 3-phase 400 V, 1400-1500 rpm or
0,75 kW, 3-phase 230 V, 1400-1500 rpm or
0,75 kW, 1-phase 230 V, 1400-1500 rpm or
1,1 kW, 3-phase 400 V, 1400-1500 rpm or
1,1 kW, 3-phase 230 V, 1400-1500 rpm or
1,1 kW, 1-phase 230 V, 1400-1500 rpm
Inlet suction check valve **
Wayne WM049971-000X
Additive liquid injection tank Wayne (START ITALIANA S.r.I. (ProGauge) gauge, 0-1pc WT010992
Pump and air separator function, LPG
Remote pump, submerged
Gas separator, non return valve and Petrolmeccanica:LP0735R01(WM058684) strainer
Vapour return line from gas separator and differential valve at volume sensor
Tokheim GPL Gas separator 950996
LPG Diff valve 943166
Non return valve 949695
Regulating function, solenoid valve, not LPG
| NEW ASC0 code | Wayne code(Datasheet in parentheses) | Wayne designation | |
| Cat no | Voltage | ||
| JV 431463-001 | 24/DC | WM045522-0001(WU007252) | SOLENOID VALVE, PROPORTIONAL,SINGLEIECEX,ALU BODY |
| JV 431463-003 | 24/DC | WM045522-0003(WU007252) | SOLENOID VALVE, PROPORTIONAL,SINGLEIECEX,ALU BODY,INMETRO,PRODUCTIONBRAZIL |
| JV 431463-002 | 24/DC | WM045522-0002(WU007252) | SOLENOID VALVE, PROPORTIONAL,SINGLEIECEX,COMPCOTE ALU BODY |
| JV 431464-001 | 24/DC | WM045523-0001(WU007253) | SOLENOID VALVE,ON/OFF,SINGLEIECEX,ALU BODY |
| JV 431465-001 | 24/DC | WM045525-0001(WU007254) | SOLENOID VALVE,PROPORTIONAL,SINGLEIECEX,BRASS BODY |
| JV 431466-001 | 24/DC | WM045526-0001(WU007255) | SOLENOID VALVE,PROPORTIONAL,DUPLEX,SINGLEEQUIPPEDRIGHT,IECEX,BRASS BODY |
| JV 431467-001 | 24/DC | WM045527-0001(WU007256) | SOLENOID VALVE,ON/OFF,DUPLEXIECEX,BRASS BODY |
| JV 430298-001 | 24/DC | WM044745-0001(WM045843) | SOLENOID VALVE,PROPORTIONAL,SINGLEIECEX,ALU BODY,CABLE LENGTH3000mm |
| JV 430298-002 | 24/DC | WM044745-0002(WM045843) | SOLENOID VALVE,PROPORTIONAL,SINGLEIECEX,ALU BODY,INMETRO,PRODUCTIONBRAZIL |
- is replaced in dispenser variant
** is left out in dispenser variant
*** is added in dispenser variant
**** is added for ATC
| JV 430298-003 | 24/DC | WM044745-0003 (WM045843) | SOLENOID VALVE, PROPORTIONAL, SINGLE IECEX, COMPCOTE ALU BODY |
| JV 430298-004 | 24/DC | WM044745-0004 (WM045843) | SOLENOID VALVE, PROPORTIONAL, SINGLE IECEX, COMPCOTE ALU BODY, CABLE LENGTH 1750 mm |
| JV 430298-005 | 24/DC | WM044745-0005 (WM045843) | SOLENOID VALVE, PROPORTIONAL, SINGLE IECEX, COMPCOTE ALU BODY, CABLE LENGTH 1750 mm |
| JV 430298-006 | 24/DC | WM044745-0006 (WM045843) | SOLENOID VALVE, PROPORTIONAL, SINGLE IECEX AND INMETRO, ALU BODY, 3 M CABLE |
| JV 430298-007 | 24/DC | WM044745-0007 (WM045843) | SOLENOID VALVE, PROPORTIONAL, SINGLE IECEX, COMPCOTE ALU BODY, 5 M CABLE |
| JV 430299-001 | 24/DC | WM044746-0001 (WM045844) | SOLENOID VALVE, ON/OFF, SINGLE IECEX, ALU BODY |
| JV 430301-001 | 24/DC | WM044747-0001 (WM045846) | SOLENOID VALVE, PROPORTIONAL, DUPLEX IECEX, BRASS BODY |
| JV 430302-001 | 24/DC | WM044750-0001 (WM045847) | SOLENOID VALVE, PROPORTIONAL, DUPLEX, SINGLE EQUIPPED RIGHT, IECEX, BRASS BODY |
| JV 430303-001 | 24/DC | WM044751-0001 (WM045850) | SOLENOID VALVE, ON/OFF, DUPLEX IECEX, BRASS BODY |
Regulating function, LPG
Differential valve, liquid pressure at least 1 bar higher than vapour pressure Manometer, between volume sensor and differential valve
Included in volume sensor LPG6000, separate after the volume sensor for GPL700. Scale interval d $ \leq $ 2 bar
Delivery function
Hose, 16 mm (5/8") , max length 10 m
Hose, 16 mm (5/8") , max length 10 m
Elaflex Conti-Slimline 16
Hose, 16 mm (5/8") , max length 10 m
Hose, 16 mm (5/8") , max length 10 m
Elaflex Conti-Slimline 16 BIO
Elaflex Conti-Slimline 16 LT
Hose, 19 mm (3/4") , max length 10 m
Goodyear EN 1360, type 3, 16 bar
Hose, 21 mm (7/8") , max length 10 m
Goodyear Hardwall Petrol Hose B.S. 3395/1989 type 3
Hose, 21 mm (7/8") , max length 6 m
Elaflex Conti-Slimline 21
Hose, 21 mm (7/8"), max length 10 m mmq = 5,01
Elaflex Conti-Slimline 21 - COAX (vapour recovery)
Hose, 21 mm (7/8") , max length 10 m
Hose, 21 mm (7/8") , max length 10 m
Elaflex Conti-Slimline 21 - COAX (vapour recovery)
Hose, 21 mm (7/8") , max length 6 m
Elaflex Conti-Slimline 21 BIO
Hose, 21 mm (7/8"), max length 10 m mmq = 5,01
Elaflex Conti-Slimline 21 LT
Elaflex Conti-Slimline 21 LT (vapour recovery)
Elaflex Conti-Slimline 21 LT (vapour recovery)
Hose, 25 mm (1") , max length 6 m
Hose, 25 mm (1") , max length 10 m, mmq = 5,01
Elaflex Conti-Slimline 25 LT
Hose, 25 mm (1") , max length 10 m
Elaflex Conti-Slimline 25 LT
Hose, 32 mm $ ( 1^{1} / 4^{*} ) $ , max length 7 m mmq = 5,01
Hose, DEF, max length 6 m, mmq = 5,0 I
Goodyear EN 1360, type 3, 16 bar
Hose, LPG, max length 6 m, mmq = 5,0 I
Hose, 25 mm (1"),max length 40 m,mmq=201 max flow rate >60 LPM
Elaflex HD 32
Elaflex LPG ID=16, max working p=25 bar
Hose, 32 mm (11/4") , max length 40 m, mmq = 201, max flow rate >60 LPM
Elaflex LPG ID=13, Elaflex LPG ID=16 or Elaflex LPG ID=19, max working p=25 bar
Nozzle 40-70 I/min
Nozzle 40-70 I/min
Nozzle 90-200 I/min
Nozzle 40-70 I/min, mmq = 5,0 I
Nozzle 90-130 I/min
Nozzle 40 I/min, mmq = 5,0 I
Nozzle, DEF
Nozzle, DEF
Elaflex ZVA slimline or OPW 21 or equivalent
Elaflex HD 25 C winded on Reel
Elaflex HD 32 C winded on Reel
Elaflex ZV 19 or equivalent
Nozzle, LPG
Elaflex ZVA slimline TMV"DRIP-STOP" or equivalent
OPW ACN or AVN or equivalent
Elaflex ZVA 25,1" or equivalent
Elaflex ZVA 32, 11/4" or equivalent
Elaflex ZVA SS AdBlue or equivalent
Elaflex ZVA AdBlue LV or equivalent
Elaflex ZVG 2, Elaflex Gasguard LPG, Boessenkool LPG, OPW LPG or equivalent
Components included for electronic calculator function, iGEM
CPU-board,0,1 or 2 (only for 4User) pc
Wayne WM001908-0001 or
Wayne WM001908-0002 or
Wayne WM001908-0005 or
Wayne WM001908-0006 or
Wayne WM001908-0011 or
Wayne WM001908-0012 or
Wayne WM001908-0015 or
Wayne WM001908-0016 or
Wayne WM069506-0001
Software CPU board, 0 or 1 pc
62D1 (Rev 12.xx) or 5A73 (Rev 15.xx) or D3CD (Rev 16.xx)
Software CPU board for dispensers
W&M checksum C10C (Rev 13.xx)
with iGEM printer, 0 or 1 pc
iGEM expansion board, 0-1 pc
Wayne WM044451-0001 or WM044451-0002
Nozzle Switch Converter (0 or 1 pcs)
Wayne WU004202-0001 or WU006752
Components included for electronic calculator function, iGEM 2
CPU-board,0,1(1 to 4 User) pc
Wayne WM044721-0001 or
Wayne WM044721-0002 or
Wayne WU020193-0001
I/O-board 0,1 pc
WM054392-0001 or WU016738-000X
Bootloader CPU board, 0 or 1 pc
F92E 4B0B (Rev 007.XXX.XXX)
Dispenser App. SW CPU board, 0 or 1 pc
2FC0 77BE (Rev 017.XXX.XXX) or F0E5 8123 (Rev 18.XXX.XXX)
iGEM2 expansion board, 0-1 pc
Wayne or DFS WM070224-000X
Ethernet communication board
Wayne WM067037-0001
Control board, Additive injection, 0-1pc
Wayne WM067598-000X
Components included for electronic function, iGEM & iGEM 2
Isolated interface communication board, 0-1 pcs Wayne WM004360
ATCL/RS422 communication board, 0-1 pcs Wayne WM001827
IFSF/LON communication board, 0-1 pcs Wayne WM008766
NPCL/SINP communication board, 0-1 pcs
Wayne WM008750
iGEM-ISB board (pulse transmitter), 1-3 pcs
TTL/CAN converter, 0-1 pc****
Wayne WU004200-0001 or WU04200-0002 or WM002450 or WM069425-000X
Software TTL/CAN converter, 0-1 pc****
Display board complete (volume, price & upd), 1-4 pcs)
Wayne WU007564-0001
W&M checksum 3FBD or 7E60 or 082A or 55AB
Display module TN neg (volume, price), 1-4 pcs
Wayne WU007562-0001 or WM065705-0001 or WM084971-0001 or WM063229 or WU015222-0001 or WM069974-0001
Display module TN neg (volume, price & upd), 1-4 pcs
POLYTRONIX, PS-030ATL-02, REV.A (WM040823-0002)
PS-039ATL-02 (WM053326-0001), upd under or PS-044ATL-02 (WM056449-0001), upd right or PS-045ATL-02 (WM056449-0002), upd left or
Display board (unit prices), 1-2 pcs
Display module TN neg (unit price), 1-12 pcs
Wayne WU007563-0001 or WM062874
POLYTRONIX,PS-030ATL-02,(WM040824-0007) or POLYTRONIX,PS031ATL-02,(WM040824 -0008) or POLYTRONIX,PS-032ATL-02,(WM040824 -0009)
Nozzle Switch Interface (0 or 1 pcs)
Wayne WU004201-0001 or WU006751-000X
Control Board iGEM printer (0,1 or 2)
Innovision 10600101100G (WR001816)
iGEM Printer (0,1 or 2)
Fujitsu FTP-629MCL374-R (WR001815)
Board; Connector, Suction pump, IEC**
Wayne WU004985
Board; Connector, dispenser, IEC***
Wayne WU004987
Board; Connector, without heating regulation Wayne WXXXXXXX-000X
and surge protection
Power supply,1 pc
Channel Well Technology, UAS150B, 150 W Max
(WM027313-0001)
Heater including thermostat, 0-1 pc Wayne WM044405 (290 W) or equivalent
(mandatory for payment terminal) approved by Wayne
Payment or Media platform tested according to OIML R117:2019 (by an OIML test laboratory) to show evidence not disturbing the pump calculator, may be installed in the dispenser, using the dispenser power supply.
Optional equipment and functions
ATC conversion device (Not for DEF, Windshield liquid and HVO)
Pre-set of volume or price ******
Volume totalizer
Blending (designation only e g 92, 95, 98 octane not 50/50, 70/30 etc)
Satellite (SAT and MASTER enable button in case of direct sales)
When using the measuring system to refill aircrafts or helicopters it is allowed to use a water elimination device according to OIML 117-1 5.8.1.2 (E 2019) between meter and hose. For safety reasons it is required to drain the device on a regular base by opening the drain valve. This has to be done by a trained service technician on the measuring system under pressure with closed nozzle. This is to ensure that the hydraulic system is always filled with fuel. The drain valve has not to be sealed.
Additive injection function for injecting additive liquids to the fuel before the meter. A tank gauge meter is required in the storage tank for additive liquid to avoid injection of air into the fuel. See requirements for central pump function.
If additive liquid is injected before the air separator, the additive is not subject to MID requirements.
Integrated equipment and functions not subject to legal metrology control
Vapour recovery system with vapour pump, regulating proportional valves and an electronic board (for example Bürkert control board, Dover Ivory power pump control or Dover Global VR) connected to the iGEM CPU board. The vapour flow is controlled by the electronic board or the iGEM CPU board. Optionally also a vapour recovery monitoring system consisting of a Vapour meter and belonging Intrinsic safe barrier may be used. Means for vapour recovery must not influence the accuracy of measurements such that the maximum permissible error is exceeded.
For DEF, heating is installed to keep the liquid from freezing. As an option a pump circulates liquid downstream the meter between deliveries, through the coax hose to a container (max volume 2,55 litre in circulation system) to avoid freezing.
- is replaced in dispenser variant **** is added in dispenser variant
** is left out in dispenser variant ***** is added for ATC
***** can be controlled through key pad on dispenser or by softkey on payment/media terminal
Rated operating conditions
Volume of liquid fuel expressed in litre. If ATC conversion device is included (not for DEF, Windshield liquid and HVO), the volume at base condition (15 $ ^{\circ} \mathrm{C} $ ) is displayed. The calculations are based on OIML R63 (API/ASTM 54B) except for petrol/ethanol mixtures where Verfahren 1 according to PTB-1.5-40520004 dated April 11, 2011, is used.
| Measurement range | |
| Maximum flowrate ($q_{\mathrm{max}}$) | 40 to 200 l/min (DEF and Windshield down to 10 l/min) |
| Minimum flowrate ($q_{\mathrm{min}}$) | $\geq$ 0,2 l/min |
| Minimum measured quantity (mmq) | 2,0 l or 5,0 l or 20 l (see specification of hoses and nozzles). As option, 5,0 l can be used on any configuration up to 5L. For LPG, 5 l only. DEF iMeter DM2-X UREA, 5,0 l only. DEF iMeter2 DM2-X DEF, 2,0 and 5,0 l. |
| Scale interval, volume display | 0,01 l |
| Maximum working pressure | 0,3 MPa, 1,8 MPa only for LPG6000 and 2,5 MPa only for GPL700 |
| Minimum working pressure | 0,12 MPa, not for LPG |
| Liquid temperature range | -30℃ to +55℃, -25℃ to +55℃ only for FAME/RME -5℃ to +35℃ only for DEF -5℃ to +35℃ only for LPG (LPG6000) -30℃ to +55℃ only for LPG (GPL700) |
| Type of liquids, volume sensor iMeter, iMeter2 and Xflo™ Viscosity range volume sensor iMeter DM2-X UREA | Petrol, kerosene, diesel, ethanol or FAME/RME/HVO 0,4 -8,0 mPas DEF (Diesel Exhaust Fluid, e.g. AdBlue) and Windshield liquid |
| volume sensor iMeter2 DM2-X DEF volume sensor LPG6000, GPL700 | DEF (Diesel Exhaust Fluid, e.g. AdBlue) LPG |
| Volume sensor flow rate range, iMeter Duplex (each meter) and Single | 0,2 to 70 l/min 4 to 70 l/min only for FAME/RME |
| iMeter DM2-X UREA | 1 to 40 l/min only for Windshield liquid |
iMeter DM2-X UREA
0,2 to 70 I/min
4 to 70 I/min only for FAME/RME
1 to 40 I/min only for Windshield liquid
2 to 40 I/min only for DEF
| iMeter2 Duplex(each meter) and Single (Flow compensating Pules Transmitter) | 0,2 to 110 l/min 0,4 to 40 l/min only for DEF |
| iMeter2 Duplex(each meter) and Single (Not flow compensating Pules Transmitter) | 1,6 to 70 l/min 0,2 to 70 l/min only for Petrol 4 to 70 l/min only for FAME/RME |
| Xflo™ Duplex(each meter) and Single | 4 to 80 l/min 4 to 70 l/min only for FAME/RME |
| LPG6000 | 5 to 50 l/min only for LPG |
| GPL700 | 10 to 50 l/min only for LPG |
| Mixture conditions,two liquids | 5% to 95% (designation only e.g.92,95,98 octane not 50/50,70/30 etc) |
The ratio between maximum and minimum flow rate should be at least 10 for single quality and 5 for blending quality, DEF, Windshield liquid and LPG.
Environments classes / influence quantities
Mechanic: class M1
Electromagnetic: class E1
Humidity: class H3
Ambient temperature limits: -40 $ ^{\circ} \mathrm{C} $ to $ +55^{\circ} \mathrm{C} $ (tested to $ +60^{\circ} \mathrm{C} $) $ -25^{\circ} \mathrm{C} $ to $ +55^{\circ} \mathrm{C} $ for DEF, Windshield liquid and LPG $ +5^{\circ} \mathrm{C} $ to $ +55^{\circ} \mathrm{C} $ for iGEM printer
Interfaces and compatibility conditions
Interface for vapour recovery systems.
Communication between fuel dispenser and other parts of a measuring system (e g POS-systems) using one of the following non-reactive interfaces is evaluated and approved. (This does not imply that the complete self-service arrangement fulfils the requirements of OIML R117):
| SW protocol | Hardware |
| DART | RS485 or RS422(CPU board) |
| LJCL | RS422(CPU board) |
| Ferranti | UKCL(CPU board) |
| IFSF | IFSF/LON communication board |
| ATCL | ATCL/RS422 communication board |
| SINP | NPCL/SINP communication board |
Most protocols do not include "unit", for example "volume @ 15 $ ^{\circ} \mathrm{C} $ ". Instead this has to be handled during set up of the station controller and dispensers. It must be possible for the measuring system to print "@ 15 $ ^{\circ} \mathrm{C} $ " on the ticket.
When the dispenser is a part of a self-service arrangement the following requirements have to be fulfilled:
The communication between the dispenser and the self-service device/devices has to be through one of the interfaces/protocols listed above.
The connection has to be verified by the installer so the metrolgical data are presented in a way, whether it is an indicated, printed or memorized result, that meets the requirements of OIML R117.
Control of the measuring tasks of the instrument in use
For dispensers with ATC the volume at base condition is displayed in normal working mode. Alternative 1: When verifying the dispenser as a "black box", the delivered volume measured in a volume standard $ V_{t\text{ref}} $ is converted to volume at base condition $ V_{15} $ using $ V_{15\text{ref}}=V_{t\text{ref}} $ x VCF where VCF, volume correction factor, is based on the quality and temperature t of the liquid. Alternative 2: When the display is showing ATC Revision and CRCs press the CRC button and the W&M-mode is active for 30 minutes (or until next push on the CRC-button) displaying $ V_{t}, $ $ V_{15}, $ temperature t, product and density. A test well (inner diameter 4 mm) can be used for the reference temperature sensor.
Amount display Product and density for primary meter
Volume display: Product and density for secondary meter
UPD: Blank
Amount display Uncompensated volume, $ V_{t} $
Volume display: Temperature for primary meter (momentary value)
UPD: Temperature for secondary meter (momentary value)
Amount display Volume at 15 $ ^{\circ} \mathrm{C} $ $ V_{1 5} $ altered with product and density for primary meter
UPD: Temperature t (mean value, volume weighted)
LPG: A thermometer well is installed upstream the meter, to be used during verification
Identification Software for iGEM
The metrological software is identified by the checksum (W&M CRC value), which can be accessed by pressing the CRC button on the iGEM & iGEM 2 calculator CPU board. (Upon this also the program CRC values and software version number are indicated at the dispenser display.)
CRC button and example of indication of software version on display
For ATC this information is altered by
| Amount display Volume display: UPD: and | W&M CRC value on board temperature module Revision of board temperature module AtC |
| Amount display Volume display: UPD: | W&M CRC value on TTL/CAN converter Revision of TTL/CAN converter Can |
Identification Software for iGEM 2
The metrological software is identified by the checksum (W&M CRC value), which can be accessed by pressing the CRC button on the iGEM 2 calculator CPU board. The software versions and CRC values are shown in the following sequence:
CRC button placement
The first indicating on the sales volume display after the CRC button is pushed is the Firmeware version of the dispenser application.
After a few seconds the display will change and show the CRC for complete program both LR and non LR is included.
After an other few seconds the display will change again and the amount and volume display will show the W&M-CRC for the LR part of the code and the UPD will show LrCrC. It will looks like this:
Example of W & M CRC
Then will the display change again and show the Bootloader Firmware version.
After that the Bootloader Firmware CRC. The UPD will show b CrC
When the display change again it will show the ATC information
Amount display: W&M CRC value on board temperature module
Volume display: Revision of board temperature module
UPD: AtC
Calibration-/adjustment procedure
Dispensers with iMeter volume sensors are calibrated/adjusted according to Wayne manual W8-461191 (English version, other languages may be available).
Dispensers with iMeter2 volume sensors are calibrated/adjusted according to Wayne manual WM064180 (English version, other languages may be available)
Dispensers with Xflo $ ^{\mathrm{TM}} $ volume sensors are calibrated/adjusted according to Wayne manual WM018625 (English version, other languages may be available).
Dispensers with LPG6000 volume sensors are calibrated/adjusted according to Wayne manual WM059375 (English version, other languages may be available).
RISE Research Institutes of Sweden AB | Certification
Dispensers with GPL700 volume sensors are calibrated/adjusted according to Wayne manual WM078856 (English version, other languages may be available)
ATC is calibrated/adjusted according to Wayne manual WM055095 (English version, other languages may be available, contact Wayne local technical support).
Sealing
The dispenser is sealed according to pictures in the following pages.
ATC: The board temperature module and temperature sensors are sealed
LPG: The volume sensor, pulser, differential valve, pressure calibration screw (if present) of the differential valve and manual valves in vapour return line are sealed (or labelled)
Name plate (aluminium or vinyl sticker) is sealed to the frame with a small "vandal proof" label.
Electromechanical totalizer is sealed if used as a basis for legal transaction.
The following components are not sealed, but protected/identified in the following way:
The pulsers are uniquely electronically identified by the calculator, including a presence check.
NOTE: When using wire through screws etc. to make seals, ensure that the wire is as short and taut as possible. The closing stamp must be placed such as it is within reach for pressing with intended pliers, without dismantle surrounding components with the help of tools. The closing stamp can be of any suitable material.
Sealing of temperature sensors on iMeter
Sealing of temperature sensors on iMeter2
Sealing of temperature sensors on LPG 6000
Sealing of temperature sensors on GPL 700
Sealing of hydraulic unit with iMeter volume sensor
Sealing of hydraulic unit with $ \mathrm{Xflo}^{\mathrm{TM}} $ volume sensor
SEALING OF PULSER AND CALIBRATION FLAPS
SECTION SHOWING PLACEMENT OF SEALING
WHERE FOR SEALING OF PULSER AND CALIBRATION
FLAPS, PLACE SEALING AS CLOSED TO SEALED PARTS AS
POSSIBLE.
Sealing of dispenser variant (central pump) with iMeter volume sensor
Sealing of dispenser variant (central pump) with $ \mathrm{Xflo}^{\mathrm{TM}} $ volume sensor
Sealing of dispenser with iMeter2 volume sensor
Sealing of LPG 6000 volume sensor
Sealing of the differential valve
Sealing between meter and differential valve against opening
Meter sealing to the dispenser frame
Sealing of the adjustment device
Sealing of the meter against opening
Sealing of the connection between meter and pulser against opening
Sealing of the differential valve
Sealing of the Meter electronic calibration/against opening
Sealing of the connection between meter and pulser against opening/removal
Sealing of GPL 700 Volume sensor
Example of sealing of iGEM 2 electronic module with electromagnetic totalizers (if used as a basis for legal transaction). Even though no electromagnetic totalizer is connected the sealing have to include the CPU-module and the sealing screw in detail 1
DETAIL A - SEALING OF CPU OPENING AND BATTERY
DETAIL B - SEALING OF CPU BOARD, I/O BOARD AND PROGRAM LOAD SWICH OPENING
DETAIL C - SEALING OF EMT OF EMT REMOVAL
DETAIL D - SEALING OF EMT CABLE CONNECTION
Example of sealing of iGEM 2 Generation 2 electronic module with electromagnetic totalizers (if used as a basis for legal transaction) Even though no electromagnetic totalizer is connected the sealing have to include the CPU-module and the sealing screw in detail A & B
Information to be borne by and to accompany the instrument
The name plate mounted on the instrument shall contain the following information:
the OIML certificate number
Space for verification mark
the name or trademark
the serial number and year of manufacture
the designation or type name (according to "Product names and designation")
Type of Liquid or Viscosity class (not for LPG, DEF, and Windshield liquid) $ ^{*} $
the temperature range of the dispensed liquid
the accuracy class
max. flow rate Qmax
min. flow rate Qmin
minimum measured quantity
max. pressure Pmax
the ambient temperature range
mechanical class
electromagnetic class
humidity class
Nominal value of the AC voltage supply
- depending of volume sensor
Further inscriptions, if necessary
Minimum measured quantity shall be inscribed on the "indicator face". If ATC is included, "@ 15 $^{\circ}C"$ shall be marked adjacent to the volume display.
Display for ATC including "@ 15 $^{\circ}C"$
Alt Display for ATC including "@ 15 $^{\circ}C"$